ProS - 0020002800 HemosIL™ ProS
Transcription
ProS - 0020002800 HemosIL™ ProS
HemosIL™ ProS - 0020002800 ENGLISH - Insert revision 11/2009 Intended use Protein S is a vitamin K-dependent cofactor for the anticoagulant and the profibrinolytic effects of activated Protein C.1 Two forms of Protein S are present in plasma: free Protein S (40%), and Protein S linked to the complement C4b-binding protein (C4BP) (60%). Only free Protein S has functional cofactor activity.2 Protein S deficiency may be hereditary or acquired. Acquired deficiency may be observed during pregnancy, oral anticoagulant therapy, oral contraceptive use, in liver disease, in newborn infants as well as in other clinical conditions.3,4,5 Deficiency of Protein S has been associated with a high risk of developing venous thromboembolism especially in young people.6,7 The ProS™ assay determines the functional activity of free Protein S by measuring the degree of prolongation of a prothrombin time in the presence of the tissue factor, phospholipids, calcium ions and activated Protein C. The Protein S activity is proportional to the prolongation of the clotting time of a Protein S deficient plasma to which diluted sample has been added. - ACL Futura® and ACL Advance Systems: 12 hours at 2-8°C, 7 days at -20°C in the original vial or 1 hour at 15°C on-board. No stirring is required. - ACL TOP ® Family: 1 hour at 15°C on-board. No stirring is required. Protein S deficient plasma - Stability after reconstitution: 4 hours at 15-25°C in the original vial or 7 days at -20°C. Protein S control plasma - Stability after reconstitution: 4 hours at 15-25°C in the original vial or 7 days at -20°C. Notes: - Frozen reagent and plasmas should be thawed at 37°C and gently mixed before use. Before freezing they can be aliquoted using plastic tubes and stoppers. Do not refreeze. - For optimal stability remove reagents from the system and store them at 2-8°C in the original vial. - The storage conditions of the reagents used to prepare the calibration curve should be the same as the reagents in use to test samples/controls (non-frozen or thawed). - On the ACL TOP Family it is recommended to calibrate with every session and to batch sample runs. Composition Instrument/test procedures Automated coagulation functional assay for the quantitative determination of free Protein S in human citrated plasma on IL Coagulation Systems. Summary and principle The ProS kit consists of: R Protein S reagent (P/N 0020002810): 3 x 3 mL vials of a lyophilized preparation containing recombinant rabbit tissue factor, synthetic phospholipids, calcium ions, activated Protein C, polybrene, buffer, stabilizers and preservatives. D Protein S deficient plasma (P/N 0020002820): 6 x 1 mL vials of a lyophilized human plasma which has been artificially depleted of Protein S. P Protein S control plasma (P/N 0020002830): 2 x 1 mL vials of a lyophilized human plasma containing a low level of Protein S. PRECAUTIONS AND WARNINGS: The material in this product was tested by FDA approved test methods and found nonreactive for Hepatitis B Surface Antigen (HBsAg), Anti-HCV and HIV antibodies. Handle as if potentially infectious.8 Avoid contact with skin and eyes (S 24/25). Do not empty into drains (S 29). Wear suitable protective clothing (S 36). This product is For in vitro Diagnostic Use. Preparation Protein S reagent: Dissolve the contents of each vial with 3 mL of CLSI (formerly NCCLS) CLRW Type water or equivalent.9 Replace the stopper and swirl gently. Make sure of the complete reconstitution of the product. Keep the reagent at 15-25°C for 30 minutes and invert to mix before use. Do not shake. Protein S deficient plasma: Dissolve the contents of each vial with 1 mL of CLSI CLRW Type water or equivalent.9 Replace the stopper and swirl gently. Make sure of the complete reconstitution of the product. Keep the plasma at 15-25°C for 30 minutes and invert to mix before use. Do not shake. Protein S control plasma: Dissolve the contents of each vial with 1 mL of CLSI CLRW Type water or equivalent.9 Replace the stopper and swirl gently. Make sure of the complete reconstitution of the product. Keep the plasma at 15-25°C for 30 minutes and invert to mix before use. Do not shake. Reagent storage and stability Unopened reagents are stable until the expiration date shown on the vial when stored at 2-8°C. Protein S reagent - Stability after reconstitution: - ACL® 100-7000 Systems: 2 days at 2-8°C, 15 days at -20°C in the original vial or 2 days at 15°C on-board. No stirring is required. - ACL 8000/9000/10000 Systems: 2 days at 2-8°C, 15 days at -20°C in the original vial or 4 hours at 15°C on-board. No stirring is required. Refer to the appropriate IL instrument’s Operator’s Manual and/or Application Manual for the complete assay procedure instructions. Specimen collection and preparation Nine parts of freshly drawn venous blood are collected into one part trisodium citrate. Refer to CLSI Document H21-A5 for further instructions on specimen collection, handling and storage.10 Additional reagents and control plasmas The following are not supplied with the kit and must be purchased separately. Americas and Pacific Rim Europe Cat. No. Cat. No. Calibration plasma 0020003700 0020003700 Normal Control 0020003120 0020003110 Low Abnormal Control 0020003220 0020003210 High Abnormal Control 0020003320 0020003310 Factor diluent 0009757600 0009757600 Quality control Normal and abnormal controls are recommended for a complete quality control program.11 Normal Control, Low Abnormal Control, High Abnormal Control and the Protein S control plasma in this kit are designed for this program. Each laboratory should establish its own mean and standard deviation and should establish a quality control program to monitor laboratory testing. Controls should be analyzed at least once every 8 hour shift in accordance to good laboratory practice. Refer to Westgard et al for identification and resolution of out-ofcontrol situtations.12 Traceability of calibrators and control materials The reported values were determined over multiple runs on IL Coagulation Systems using a specific lot of reagent and against a Calibration Plasma House Standard which is traceable to the current International Protein S Standard, identified in the acceptance range table. Results Limitations/interfering substances ProS results on the ACL Family and ACL Futura/ACL Advance Systems are not affected by heparin (UF heparin or LMW heparin) up to 2 U/mL, bilirubin up to 15 mg/dL, hemoglobin up to 200 mg/dL and triglycerides up to 1000 mg/dL. ProS results on the ACL TOP Family are not affected by heparin (UF) up to 1.6 U/mL, heparin (LMW) up to 1.8 U/mL, bilirubin up to 15 mg/dL, hemoglobin up to 200 mg/dL and triglycerides up to 1000 mg/dL. Hemolyzed and turbid samples should not be assayed. ProS assay results may be affected (decrease in PS activity) by the presence of both activated Factor VII (F VIIa) 13 and Factor V Leiden mutation (APC-R), above all in patients homozygous for this genetic defect. 14 To avoid a possible influence of APC-R on the actual Protein S values, patient samples with results outside the normal range should be manually diluted 1:2 with Protein S deficient plasma and reassayed. The printed results must be multiplied by 2 to correct for the dilution. Expected values A normal range study was performed using the ProS kit on IL Coagulation Systems. System N % PS Activity ACL Family 78 63 - 135 ACL Futura/ACL Advance 80 67 - 132 ACL TOP Family 49 76 - 135 Note: Age and hormonal status may affect the normal range for females.15 Ranges were calculated as recommended by CLSI document C28-A. 16 These results were obtained using a specific lot of reagent. Due to many variables which may affect results, each laboratory should establish its own normal range. Performance characteristics Precision: Within run and between run precision was assessed over multiple runs. ACL Family Mean (% PS) CV% (Within run) CV% (Total) Normal Control 91.1 3.2 7.0 Protein S Control 52.5 8.2 10.2 ACL Futura/ACL Advance Mean (% PS) CV% (Within run) CV% (Total) Normal Control 90.5 3.9 5.1 Protein S Control 49.1 6.6 7.6 ACL TOP Family Mean (% PS) CV% (Within run) CV% (Total) Normal Control 95.6 3.1 5.6 Protein S Control 30.3 3.1 11.2 Correlation: System ACL Family ACL Futura/ACL Advance ACL TOP Family slope 1.01 1.02 1.061 intercept -5.883 -4.890 -4.832 r Reference method 0.982 IL Clotting Protein S on ACL 0.984 IL Clotting Protein S on ACL 0.985 IL ProS on ACL Futura/ ACL Advance The precision and correlation results were obtained using specific lots of reagents and controls. Linearity: System IL Coagulation System 10 - 150 (% PS activity) ProS results may be reported in seconds and activity (%). Refer to the instrument’s Operator’s manual for additional information. ProS - 0020002800 DEUTSCH - Packungsbeilage Version 11/2009 Verwendung Zur automatisierten Bestimmung des funktionellen Protein S in menschlichem Plasma auf IL Analysensystemen. Testprinzip Protein S ist ein Vitamin K-abhängiger Kofaktor für die antikoagulatorischen und profibrinolytischen Effekte von aktiviertem Protein C.1 Es liegt im Plasma in zwei Formen vor: zu 40% als Freies Protein S und zu 60 % in gebundener Form an C4bBindeprotein (Komplementfaktor). Die beiden Formen stehen in einem dynamischen Gleichgewicht. Nur das freie Protein S ist biologisch aktiv.2 Ein Protein S Mangel kann sowohl angeboren als auch erworben sein. Erworbene Mangelzustände können während der Schwangerschaft, unter oraler Antikoagulantientherapie, bei Einnahme oraler Kontrazeptiva, bei Lebererkrankungen, bei Neugeborenen und bei anderen klinischen Zuständen auftreten.3,4,5 Protein S-Mangel ist als Risikofaktor bekannt, der die Bildung von venösen Thrombosen besonders bei jungen Patienten begünstigt.6,7 Beim ProS wird die funktionelle Aktivität des Freien Protein S, als Kofaktor des aktivierten Protein C, über die Verlängerung der Prothrombinzeit in Anwesenheit von Gewebefaktor, Phospholipiden, Kalziumionen und aktiviertem Protein C bestimmt. Dadurch liegen alle übrigen Gerinnungsfaktoren im Überschuss vor und die Gerinnungszeit Bestimmung ist von der Menge an Protein S im Patientenplasma abhängig. Inhalt Die ProS Packung enthält: R Protein S reagent (Best.-Nr. 0020002810): 3 Flaschen x 3 mL lyophilisiertes Material mit rekombinantem Kaninchen-Gewebefaktor, synthetischen Phospholipiden, Kalziumionen, aktiviertem Protein C, Polybren, Puffer, Stabilisatoren und Konservierungsmitteln. D Protein S deficient plasma (Best.-Nr. 0020002820): 6 Flaschen x 1 mL lyophilisiertes Humanplasma, hergestellt durch spezifische Immunabsorption. P Protein S control plasma (Best.-Nr. 0020002830): 2 Flaschen x 1 mL lyophilisiertes Humanplasma mit einem niedrigen Protein S Gehalt. VORSICHTSMAßNAHMEN UND WARNHINWEISE: Das verwendete Material wurde mit FDA anerkannten Testmethoden auf HIVAntikörper, Hepatitis-B-Antigen und HCV-Antigen geprüft. Bitte beachten Sie die Bestimmungen zum Umgang mit potentiell infektiösen Materialien.8 Berührung mit den Augen und der Haut vermeiden (S 24/25). Nicht in die Kanalisation gelangen lassen (S 29). Bei der Arbeit geeignete Schutzkleidung tragen (S 36). Dieses Reagenz ist ausschließlich für die in vitro Diagnostik bestimmt. Herstellung lösen.9 Protein S reagent: Inhalt der Flasche mit 3 mL Aqua dest. Dann die Flasche verschließen und durch leichtes Schwenken das Lyophilisat lösen. Nach vollständiger Rekonstitution wird das Reagenz 30 Minuten bei 15-25°C inkubiert und unter vorsichtigem Schwenken erneut gemischt. Nicht schütteln. Protein S deficient plasma: Inhalt der Flasche mit 1 mL Aqua dest. lösen.9 Dann die Flasche verschließen und durch leichtes Schwenken das Lyophilisat lösen. Nach vollständiger Rekonstitution wird das Reagenz 30 Minuten bei 15-25°C inkubiert und unter vorsichtigem Schwenken erneut gemischt. Nicht schütteln. Protein S control plasma: Inhalt der Flasche mit 1 mL Aqua dest. lösen.9 Dann die Flasche verschließen und durch leichtes Schwenken das Lyophilisat lösen. Nach vollständiger Rekonstitution wird das Reagenz 30 Minuten bei 15-25°C inkubiert und unter vorsichtigem Schwenken erneut gemischt. Nicht schütteln. Lagerung und Haltbarkeit Die ungeöffneten Reagenzien sind bei Lagerung zwischen 2-8°C bis zu dem auf dem Etikett angegebenen Verfallsdatum haltbar. Protein S reagent Haltbarkeit nach Rekonstitution in ACL Systemen (100-7000): - bei 2-8°C in der Originalflasche: 2 Tage - bei -20°C in der Originalflasche: 15 Tage - bei 15°C im Gerät: 2 Tage Muss nicht gerührt werden. Haltbarkeit nach Rekonstitution in ACL 8000/9000/10000 Systemen: - bei 2-8°C in der Originalflasche: 2 Tage - bei -20°C in der Originalflasche: 15 Tage - bei 15°C im Gerät: 4 Stunden Muss nicht gerührt werden. Haltbarkeit nach Rekonstitution in ACL Futura und ACL Advance Systemen: - bei 2-8°C in der Originalflasche: 12 Stunden - bei -20°C in der Originalflasche: 7 Tage - bei 15°C im Gerät: 1 Stunde Muss nicht gerührt werden. - bei 15°C in Systemen der ACL TOP® Familie: 1 Stunde im Gerät. Kein Rühren notwendig. Protein S deficient plasma - Haltbarkeit nach Rekonstitution: - bei 15-25°C in der Originalflasche: 4 Stunden - bei -20°C: 7 Tage Protein S control plasma - Haltbarkeit nach Rekonstitution: - bei 15-25°C in der Originalflasche: 4 Stunden - bei -20°C: 7 Tage Hinweis: - Tiefgefrorenes Reagenz und Plasma sollte bei 37°C aufgetaut und vor dem Gebrauch vorsichtig geschwenkt werden. Das Produkt kann nach der Rekonstitution in Plastikgefäßen mit Deckel aliquotiert eingefroren werden. Alle Komponenten des Protein S Kits dürfen nur einmal eingefroren werden. - Für eine optimale Haltbarkeit sollten die Reagenzien nach dem Gebrauch aus dem Gerät entnommen und im Kühlschrank bei 2-8°C in der Originalflasche aufbewahrt werden. - Die Lagerungsbedingungen der Reagenzien, mit denen die Kalibrationskurve erstellt wurde, sollten für die Durchführung von Patienten/Kontrollen gleich sein (entweder nicht gefroren oder aufgetaut). - Auf Systemen der ACL TOP Familie wird empfohlen mit jedem Probenlauf zu kalibrieren und die Proben in einem Arbeitsgang abzuarbeiten. Ergebnisse Bestimmungsansatz Referenzbereiche Die ausführliche Beschreibung des Bestimmungsansatzes ist dem GeräteBedienerhandbuch und dem Applikationshandbuch zu entnehmen. Mit dem ProS Kit wurde auf IL Analysensystemen eine Studie zur Bestimmung des Normalbereichs durchgeführt. System N % PS Aktivität ACL Familie 78 63 - 135 ACL Futura/ACL Advance 80 67 - 132 ACL TOP Familie 49 76 - 135 Hinweis: Der Normalbereich für Frauen kann vom Alter und hormonellen Status beeinflusst werden.15 Die Bereiche wurden entsprechend dem CLSI Dokument C28-A berechnet.16 Die Ergebnisse sind mit einer spezifischen Reagenziencharge ermittelt worden. Aufgrund verschiedener Variablen, die die Testergebnisse beeinflussen können, wird empfohlen, dass jedes Labor seinen eigenen Normalbereich für das ProS ermittelt. Präzision Die Präzision im Lauf und zwischen Läufen wurde bei mehreren Läufen bestimmt. Probenmaterial - Antikoaguliertes Plasma 9 Teile frisches venöses Blut und 1 Teil Trinatriumcitratlösung werden sorgfältig in einem silikonisierten Glasröhrchen gemischt. Hinweise zur Aufbereitung des Blutes sind den Empfehlungen des Deutschen Instituts für Normung - DIN 58 905 - oder dem CLSI Dokument H21-A5 zu entnehmen.10 Zusätzliche Reagenzien und Kontrollplasmen Die folgenden Reagenzien sind nicht in der IL ProS Packung enthalten und müssen zusätzlich bestellt werden: Amerikan. und Pazifischer Raum Europa Art. Nr. Art. Nr. Kalibrationsplasma 0020003700 0020003700 Normal-Kontroll-Plasma 0020003120 0020003110 Low Abnormal-Kontroll-Plasma 0020003220 0020003210 High Abnormal-Kontroll-Plasma 0020003320 0020003310 Faktor Diluent 0009757600 0009757600 Qualitätskontrolle Es ist gängige Laborpraxis, die Qualität der Analyse mit Kontrollmaterialien im normalen sowie im pathologischen Bereich zu überprüfen.11 Es wird empfohlen, als Kontrollmaterial IL Normal-Kontroll-Plasma, Low Abnormal-Kontroll-Plasma und High Abnormal-Kontroll-Plasma sowie das Protein S Kontrollplasma zu verwenden. Die Bereiche sind der jeweiligen Packungsbeilage zu entnehmen. Jedes Labor sollte jedoch für das Qualitätskontroll-Programm seinen eigenen Mittelwert ± 2 Standardabweichung (s) ermitteln. Spätestens nach jeweils 8 Stunden sollte eine Qualitätskontrolle durchgeführt werden. Algorithmen zur Beurteilung der Qualitätskontrollergebnisse siehe z.B. Westgard et al.12 Siehe auch “Richtlinien der Bundesärztekammer zur Qualitätssicherung quantitativer laboratoriumsmedizinischer Untersuchungen“ vom 24.8.2001. Standardisierung des Kalibrators und der Kontrollmaterialien Die angegebenen Referenzwerte wurden bei mehreren Läufen an IL Analysensystemen mit spezifischen Reagenzienchargen und gegen ein Haus-Standard-Kalibrationsplasma ermittelt. Dieses ist abgeglichen auf den aktuellen Internationalen Standard für Protein S (siehe abschnitt Referenzbereiche). ProS - 0020002800 Die Ergebnisse der Patientenplasmen werden automatisch in folgenden Einheiten angegeben: - Sekunden - Aktivität in % ProS Weitere Informationen sind dem Bedienerhandbuch zu entnehmen. Einschränkungen ProS Ergebnisse auf ACL und ACL Futura/Advance Analysensystemen werden durch Konzentrationen an Heparin (unfraktionierte oder niedermolekulare Heparine) bis zu 2 U/mL, Bilirubin bis zu 15 mg/dL, Hämoglobin bis zu 200 mg/dL und Triglyceride bis 1000 mg/dL nicht beeinflusst. ProS Ergebnisse auf Systemen der ACL TOP Familie werden durch Konzentrationen an Heparin (unfraktioniert) bis zu 1.6 U/mL, LMW-Heparin bis zu 1.8 U/mL, Bilirubin bis zu 15 mg/dL, Hämoglobin bis zu 200 mg/dL und Triglyceride bis zu 1000 mg/dL nicht beeinflusst. Hämolytische und trübe (z.B. ikterische/lipämische) Proben sollten nicht analysiert werden. Die Bestimmung von ProS kann durch die Anwesenheit von aktiviertem Faktor VII (F VIIa)13 und Faktor V Leiden Mutation (APC-R)14 beeinflusst werden (scheinbare Verringerung der PS Aktivität), vor allem bei Patienten mit homozygoter Ausprägung dieser Genmutation. Um einen möglichen Einfluß, sowohl von APC-Resistance als auch F VIIa auf die tatsächliche ProS Aktivität der gemessenen Proben auszuschließen, sollten die Proben, deren Werte außerhalb des Normalbereichs liegen, manuell 1:2 mit Protein SMangelplasma verdünnt und nochmals gemessen werden. Anschließend müssen die erhaltenen Ergebnisse, entsprechend der eingesetzten Verdünnung, mit 2 multipliziert werden. ACL Familie Mittelwert (% PS) VK % (im Lauf) VK % (Tag zu Tag) Normal-Kontroll-Plasma Protein S Kontrollplasma 91,1 52,5 3,2 8,2 7,0 10,2 ACL Futura/ACL Advance Mittelwert (% PS) VK % (im Lauf) VK % (Tag zu Tag) Normal-Kontroll-Plasma Protein S Kontrollplasma 90,5 49,1 3,9 6,6 5,1 7,6 Mittelwert (% PS) VK % (im Lauf) VK % (Tag zu Tag) ACL TOP Familie Normal-Kontroll-Plasma 95,6 Protein S Kontrollplasma 30,3 Korrelation: System Steigung Ordinatenabschnitt ACL Familie 1,01 -5,883 ACL Futura 1,02 -4,890 ACL Advance ACL TOP Familie 1,061 -4,832 3,1 3,1 5,6 11,2 r Vergleichsmethode 0,982 IL Clotting Protein S am ACL 0,984 IL Clotting Protein S am ACL 0,985 IL Clotting Protein S am ACL Futura/ACL Advance Die Ergebnisse sind mit spezifischen Reagenzien- und Kontrollchargen ermittelt worden. Linearität: System IL Gerinnungssysteme 10 - 150 (% Aktivität PS) ESPAÑOL - Revisión Prospecto 11/2009 Aplicación Técnica coagulativa automatizada funcional para la determinación cuantitativa de la actividad de la Proteína S libre en plasma humano citratado en los Sistemas de Coagulación IL. Principio La Proteína S es un cofactor vitamino-K dependiente que interviene en los procesos anticoagulantes y profibrinolíticos de la Proteína C activada.1 La Proteína S se presenta en el plasma en dos formas: Proteína S Libre (40%) y Proteína S ligada al complemento C4-Binding Protein (C4BP) (60%). Sólo la Proteína S Libre tiene una actividad funcional como cofactor.2 Las deficiencias en Proteína S pueden ser de origen hereditario o adquiridas. Las deficiencias adquiridas se pueden observar durante el embarazo, terapia anticoagulante oral, contraceptivos orales, enfermedades hepáticas, en neonatos, así como en diversas condiciones clínicas.3,4,5 Deficiencias en Proteína S se han asociado a un alto riesgo de desarrollar trombo embolismo venoso, especialmente en gente joven. 6,7 El kit ProS™ determina la actividad funcional de la Proteína S libre, midiendo el grado de prolongación del Tiempo de Protrombina (TP) en presencia de factor tisular, fosfolípidos, iones calcio y Proteína C activada. La actividad de la Proteína S es proporcional a la prolongación del tiempo de coagulación del plasma deficiente en Proteína S al cual se le ha añadido muestra de paciente diluida. Composición El kit ProS consta de: R Protein S reagent (Núm. Cat. 0020002810): 3 x 3 mL viales de una preparación liofilizada que contiene factor tisular recombinante de conejo, fosfolípidos sintéticos, iones calcio, proteína C activada, polibrene, tampón, estabilizantes, y conservantes. D Protein S deficient plasma (Núm. Cat. 0020002820): 6 x 1 mL viales de plasma humano liofilizado del que se le ha eliminado artificialmente la Proteína S. P Protein S control plasma (Núm. Cat. 0020002830): 2 x 1 mL viales de plasma humano liofilizado que contiene una concentración baja de Proteina S. MEDIDAS DE PRECAUCIÓN Y ADVERTENCIAS El material usado en este producto ha sido verificado por métodos aprobados por la FDA y encontrado no reactivo al Antígeno de Superficie de la Hepatitis B (HBsAg), Anti-HCV y anticuerpos HIV. Manejar con precaución como si fuese potencialmente infeccioso.8 Evítese el contacto con los ojos y la piel (S 24/25). No tirar los residuos por el desagüe (S 29). Úsese indumentaria protectora adecuada (S 36). Este producto es para diagnóstico in vitro. Preparación Protein S reagent: Disolver el contenido de cada vial con 3 mL de agua mL de agua tipo CLRW o equivalente según CLSI (antes NCCLS).9 Cerrar el vial y homogeneizar suavemente. Asegurarse de la completa disolución del producto. Mantener el reactivo entre 15 y 25°C durante 30 minutos. Mezclar por inversión del vial antes de su uso. No agitar Protein S deficient plasma: Disolver el contenido de cada vial con 1 mL de agua tipo CLRW o equivalente según CLSI.9 Cerrar el vial y homogeneizar suavemente. Asegurarse de la completa disolución del producto. Mantener el reactivo entre 15 y 25°C durante 30 minutos. Mezclar por inversión del vial antes de su uso. No agitar. Protein S control plasma: Disolver el contenido de cada vial con 1 mL de agua tipo CLRW o equivalente según CLSI.9 Cerrar el vial y homogeneizar suavemente. Asegurarse de la completa disolución del producto. Mantener el reactivo entre 15 y 25°C durante 30 minutos. Mezclar por inversión del vial antes de su uso. No agitar. Conservación y estabilidad de los reactivos Los viales que no hayan sido abiertos y se hayan conservado a 2-8°C son estables hasta la fecha de caducidad indicada en la etiqueta. Protein S reagent - Estabilidad después de su reconstitución: - Sistemas ACL (100-7000): 2 días a 2-8°C, 15 días a -20°C en el vial original ó 2 días a 15°C en máquina. No requiere agitación. - Sistemas ACL 8000/9000/10000: 2 días a 2-8°C, 15 días a -20°C en el vial original ó 4 horas a 15°C en máquina. No requiere agitación. - Sistemas ACL Futura y ACL Advance: 12 horas a 2-8°C, 7 días a -20°C en el vial original ó 1 hora a 15°C en máquina. No requiere agitación. - Familia ACL TOP: 1 hora a 15 ºC en máquina. No requiere agitación. Protein S deficient plasma - Estabilidad después de la reconstitución: 4 horas a 15-25°C en el vial original ó 7 días a -20°C. Protein S control plasma - Estabilidad después de la reconstitución: 4 horas a 15-25°C en el vial original ó 7 días a -20°C. Notas: - Los plasmas y los reactivos congelados se deberán descongelar a 37°C. Homogeneizar suavemente antes de su uso. Antes de congelar los productos deberán ser repartidos en partes alícuotas usando tubos de plástico con tapón. No recongelar. - Para obtener una estabilidad óptima de los reactivos, sugerimos que una vez acabado el trabajo retire los reactivos del analizador y los conserve a 2-8°C en su vial original. - Las condiciones de almacenamiento de los reactivos (sin congelar/descongelados) deben ser las mismas tanto para realizar la curva de calibración como para realizar las analíticas de las muestras/controles. - En los analizadores de la Familia ACL TOP se recomienda calibrar en cada sesión y a continuación procesar las muestras. Método de ensayo Seguir las instrucciones de la técnica de acuerdo al Manual de Operador y/o al Manual de Aplicaciones de los instrumentos IL. Recolección y preparación de las muestras Recoger nueve partes de sangre recién extraída por punción venosa sobre una parte de anticoagulante citrato trisódico. Para la recolección, manipulación y conservación de la muestra seguir las recomendaciones del Documento H21-A5 de la NCCLS.10 Reactivos y plasmas de control adicionales Los siguientes productos no se suministran con el kit y deberán pedirse por separado. América y Área del Pacífico Europa Núm. Cat. Núm. Cat. Plasma de Calibración 0020003700 0020003700 Control Normal 0020003120 0020003110 Control Anormal Bajo 0020003220 0020003210 Control Anormal Alto 0020003320 0020003310 Diluyente de factores 0009757600 0009757600 Control de Calidad Para realizar un programa completo de Control de Calidad, se recomienda el uso de controles normal y patológico.11 Los controles de IL adecuados para este programa son el Control Normal, el Control Anormal Bajo, el Control Anormal Alto y el Plasma Control Proteína S incluido en el kit. Cada laboratorio debe establecer su propia media y desviación estándar y debe realizar un programa de Control de Calidad para monitorizar sus resultados. Los controles deben ser analizados una vez por cada turno de 8 horas, de acuerdo con la Normativa de Buenas Prácticas de Laboratorio. Consúltese el Manual del Operador para obtener más información. Consúltese la publicación de Westgard y col. para la identificación y resolución de situaciones anormales del Control de Calidad.12 Trazabilidad de los calibradores y controles Los valores indicados en la ficha técnica fueron obtenidos a través de múltiples determinaciones en los sistemas de coagulación de IL usando un lote específico de reactivo y un Estándar Interno de Plasma Calibrador. Dicho Estándar está valorado respecto al actual Estándar Internacional de Proteína S el cual está indicado en la tabla de valores limites aceptados. Los resultados de ProS se expresan en segundos y actividad (%). Referirse al Manual del Operador para mas información. Limitaciones/interferencias Los resultados de ProS en los sistemas de la Familia ACL y de la Familia ACL Futura/ACL Advance no se ven afectados por concentraciones de Heparina (heparina estándar y de bajo peso molecular) hasta 2 UI/mL, bilirrubina hasta 15 mg/dL, hemoglobina hasta 200 mg/dL y triglicéridos hasta 1000 mg/dL. Los resultados de la ProS en la Familia ACL TOP no son alterados por concentraciones de heparina no fraccionada de hasta 1.6 UI/ML, de heparina de Bajo Peso Molecular de hasta 1.8 UI/mL, niveles de bilirrubina de hasta 15 mg/dL, niveles de hemoglobina de hasta 200 mg/dL y niveles de triglicéridos de hasta 1000 mg/dL. Plasmas hemolizados o turbios no deberían ser analizados. Los resultados de ProS pueden verse afectados (disminuyendo la actividad PS) tanto por la presencia en la muestra del Factor VII activado (Factor VIIa)13 como por la mutación del Factor V (Factor V Leiden/APC-R)14, sobre todo en pacientes homozigóticos para esta mutación genética. Para evitar una posible influencia de la APC-R en los valores de la actividad ProS, las muestras que informen resultados fuera del rango normal deben diluirse manualmente 1:2 con el plasma deficiente en Proteína S y reanalizarse. Los resultados obtenidos deben multiplicarse por el factor de dilución 2. Valores esperados Se ha realizado un estudio del rango de normalidad utilizando el kit ProS en los sistemas de coagulación de IL. Sistema N PS (Actividad %) Familia ACL 78 63 - 135 ACL Futura/ACL Advance 80 67 - 132 Familia ACL TOP 49 76 - 135 *Nota: la edad y estado hormonal pueden afectar el rango normal en mujeres. 15 Los rangos fueron calculados siguiendo las recomendaciones del documento C28-A de la CLSI.16 Estos resultados se obtuvieron utilizando un lote específico de reactivo. Debido a las variables que pueden afectar los resultados, cada laboratorio debe establecer su propio rango de normalidad de ProS. Características técnicas Precisión: La precisión intra-serie e inter-serie se ha determinado después de realizar múltiples series de análisis. Familia ACL Media (% PS) CV% (intra-serie) CV% (Total) Control Normal 91,1 3,2 7,0 Control Proteína S 52,5 8,2 10,2 ACL Futura/ACL Advance Media (% PS) CV% (intra-serie) CV% (Total) Control Normal 90,5 3,9 5,1 Control Proteína S 49,1 6,6 7,6 Familia ACL TOP Media (% PS) CV% (intra-serie) CV% (Total) Control Normal 95,6 3,1 5,6 Control Proteína S 30,3 3,1 11,2 Correlación: Sistema Pendiente Intersección r Método de referencia Familia ACL 1,01 -5,883 0,982 Proteína S IL Coagulativa en ACL ACL Futura 1,02 -4,890 0,984 Proteína S IL Coagulativa en ACL ACL Advance Familia ACL TOP 1,061 -4,832 0,985 Proteína S IL Coagulativa en ACL Futura/ACL Advance La precisión y los datos de correlación se han obtenido usando lotes específicos de reactivos y de controles. Linealidad: Sistema Sistemas Coagulación IL 10 - 150 (% Actividad PS) Resultados Bibliography / Literatur / Bibliografía / Bibliographie / Bibliografia /Bibliografia 1. Walker FJ. Regulation of Activated Protein C by a New Protein: A Possible Function for Bovine Protein S, J. Biol. Chem. 1980; 255: 5521-5524. 2. Suzuki K, Nishioka J. Plasma Protein S Activity using Protac, a Snake Venom Derived Activator of Protein C, Thromb. Res. 1988; 49: 214-251. 3. Mannucci PM, Valsecchi C, Krachmalnicoff A, Faioni EM, Tripodi A. Familial Dysfunction of Protein S, Thromb. Haemost. 1989; 62: 763-766. 4. D’Angelo A, Viganò D’Angelo S, Esmon CT, Comp PC. Acquired Deficiencies of Protein S. Protein S Activity during Oral Anticoagulation, Liver Disease and Disseminated Intravascular Coagulation, J. Clin. Inves. 1988; 81: 1445-1454. 5. Comp PC, Esmon CT. Recurrent Venous Thromboembolism in Patients with Partial Deficiency of Protein S, N. Engl. J. Med. 1984; 311: 1522-1528. 6. Maccaferri M, Legnani C, Preda L, Palareti G. Protein S Activity in Patients with Heredofamilial protein S Deficiency and in patients with Juvenile venous Thrombosis, Thromb. Res. 1991; 64: 647-658. 7. Faioni EM, Valsecchi C, Palla A, Taioli E, Razzari C, Mannucci PM. Free Protein S Deficiency Is a Risk Factor for Venous Thrombosis. Thromb. Haemost. 1997; 78: 1343-1346. 8. Richmond JY, McKinney RW eds. Biosafety in Microbiological and Biomedical Laboratories, U.S. Dept. of Health and Human Services, Public Health Service, 4th Edition, 1999. 9. Clinical and Laboratory Standards Institute/CLSI. Preparation and testing of reagent Water in the Clinical Laboratory, Third Edition, CLSI/NCCLS Document C3-A4; Vol. 26 No. 22. 10. Clinical and Laboratory Standards Institute/CLSI. Collection, Transport and Processing of Blood Specimens for Testing Plasma-Based Coagulation and Molecular Hemostasis Assays; Approved Guideline - Fifth Edition, CLSI/NCCLS Document H21-A5; Vol. 28 No. 5. 11. Zucker S, Cathey MH, West B. Preparation of Quality Control Specimens for Coagulation, Am. J. Clin. Pathol. 1970; 53: 924-927. 12. Westgard JO, Barry PL. Cost-Effective Quality Control: Managing the Quality and Productivity of Analytical Processes, AACC Press 1986. 13. Preda L, Tripodi A, Valsecchi C, Lombardi A, Finotto E, Mannucci PM. A Prothrombin Time-Based Functional Assay of Protein S, Thromb. Res. 1990; 60: 19-32. 14. Faioni EM, Franchi F, Asti D, Sacchi E, Bernardi F, Mannucci PM. Resistance to Activated Protein C in Nine Thrombophilic Families: Interference in a Protein S Functional Assay, Thromb. Haemostas. 1993; 70: 1067-1071. 15. Liberti G, Bertina RM, Rosendaal FR. Hormonal State rather than Age Influences Cut-off Values of Protein S: Reevaluation of the Thrombotic Risk Associated with Protein S Deficiency, Thromb. Haemost. 1999; 82: 1093-1096. 16. Clinical and Laboratory Standards Institute/CLSI. How to define and Determine Reference Intervals in the Clinical Laboratory; Third Edition, CLSI/NCCLS Document C28-A3; Vol. 28 No. 30. ACCEPTANCE RANGE ACL Family and ACL Futura/ACL Advance ACL TOP Family % % P WHO STD ACL, ACL Futura and ACL TOP are trademarks of Instrumentation Laboratory. ©2002 Instrumentation Laboratory Symbols used / Verwendete Symbole / Símbolos utilizados / Symboles utilisés / Simboli impiegati / Símbolos utilizados ACCEPTANCE RANGE WHO STD In vitro diagnostic medical device In-vitro Diagnostikum De uso diagnóstico in vitro Dispositif mèdical de diagnostic in vitro Per uso diagnostico in vitro Dispositivo médico para utilização em diagnóstico in vitro Acceptance range / Referenzbereich / Rango esperado / Valeur usuelles / Intervallo di accettabilità / Intervalo de aceitação WHO Standard Code / WHO Standard Code / Código del Estándar OMS / Code standard OMS / Codice OMS / Código Padrão WHO Batch code Chargen-Bezeichnung Identificación número de lote Désignation du lot Numero del lotto Número de lote Use by Verwendbar bis Caducidad Utilisable jusqu’à Da utilizzare prima del Data límite de utilização Temperature limitation Festgelegte Temperatur Temperatura de Almacenamiento Températures limites de conservation Limiti di temperatura Límite de temperatura Instrumentation Laboratory Company - Bedford, MA 01730-2443 (USA) Instrumentation Laboratory SpA - V.le Monza 338 - 20128 Milano (Italy) Consult instructions for use Beilage beachten Consultar la metódica Lire le mode d’emploi Vedere istruzioni per l’uso Consultar as instruções de utilização Control Kontrollen Control Contrôle Controllo Controlo Biological risks Biologisches Risiko Riesgo biológico Risque biologique Rischio biologico Risco biológico Manufacturer Hergestellt von Fabricado por Fabricant Prodotto da Fabricado por Authorised representative Bevollmächtigter Representante autorizado Mandataire Rappresentanza autorizzata Representante autorizado 303049 R9 11/2009 HemosIL™ ProS - 0020002800 FRANÇAIS - Révision Notice 11/2009 Utilisation Dosage automatisé de la Protéine S fonctionnelle par méthode chronométrique. Ce dosage permet une détermination quantitative de la Protéine S libre dans les plasmas humains citratés sur les appareils de coagulation IL. Principe La Protéine S, protéine vitamine K dépendante, est le cofacteur de la Protéine C activée qui possède un effet anticoagulant et profibrinolytique.1 Deux formes sont présentes dans le plasma : la Protéine S libre (40 %), et la Protéine S liée à la C4b-binding Protein (60 %). Seule la Protéine S libre est biologiquement active.2 Les déficits en Protéine S peuvent être héréditaires ou acquis. Les déficits acquis peuvent être observés pendant la grossesse, les traitements anticoagulants oraux, l’utilisation de contraceptifs oraux, au cours les maladies hépatiques chez l’enfant nouveau-né ainsi que dans d’autres conditions cliniques.3,4,5 Les déficits en Protéine S sont associés à un risque élevé de développement d’épisodes thromboemboliques veineux, particulièrement chez le sujet jeune.6,7 Le dosage ProS™ détermine l’activité fonctionnelle de la Protéine S libre en mesurant le degré d’allongement d’un Temps de Quick (TQ) en présence de facteur tissulaire, de phospholipides, de calcium et de Protéine C activée. L’activité en Protéine S est proportionnelle à l’allongement du temps de coagulation d’un plasma déficient en Protéine S auquel le spécimen dilué a été ajouté. Composition La trousse ProS contient: R Protein S reagent (Réf. 0020002810): 3 flacons de 3 ml d’une préparation lyophilisée contenant du facteur tissulaire recombinant de lapin, des phospholipides synthétiques, du calcium, de la protéine C activée, du polybrène, du tampon, des stabilisants et conservateurs. D Protein S deficient plasma (Réf. 0020002820): 6 flacons de 1 ml de plasma humain lyophilisé, artificiellement déplété en Protéine S. P Protein S control plasma (Réf. 0020002830): 2 flacons de 1 ml de plasma humain lyophilisé contenant un faible taux de Protéine S. PRÉCAUTIONS D’EMPLOI ET MISES EN GARDE: Ces réactifs contiennent des produits d’origine humaine. Ils ont été trouvés négatifs pour les anticorps anti VIH 1/2, anti VHC et l’antigène de surface de l’hépatite B (AgHBs), en utilisant des trousses de dépistage de 3ème génération. Cependant, aucune technique ne permettant d’assurer l’absence totale du virus HIV ou de l’hépatite B ou de tout autre agent infectieux, ces réactifs sont à manipuler avec les précautions d’usage.8 Éviter le contact avec la peau et les yeux (S 24/25). Ne pas jeter les résidus à l’égout (S 29). Porter un vêtement de protection approprié (S 36). Ces produits sont à usage diagnostic in vitro. Notes: - Les réactifs et plasmas congelés pourront être décongelés à 37°C et mélangés doucement avant utilisation. Avant congélation, ils peuvent être aliquotés en utilisant des tubes en plastique et des bouchons. Ne pas recongeler. - Pour une stabilité optimale, conserver les réactifs à 2-8°C dans les flacons d’origine entre chaque utilisation. - Les conditions de conservation des réactifs utilisés pour générer la courbe d’étalonnage doivent être identiques à celles des réactifs servant à réaliser les analyses des échantillons de patients et les contrôles (aucune congélation / décongélation). - Sur les analyseurs ACL TOP, nous vous recommandons d’analyser les échantillons par série et de calibrer à chaque série. Procédures de test/instrument Se référer au mode d’emploi de l’instrument IL approprié et/ou au manuel d’application pour avoir des informations complètes sur les procédures de dosage. Recueil des spécimens et préparation 9 parts de sang fraîchement prélevé sont collectées avec 1 part de citrate trisodique. Se référer au document CLSI H21-A5 ou au document GEHT (STV, numéro spécial, 1-40, 1998) pour plus d’informations sur le prélèvement des échantillons, leur manipulation et leur stockage.10 Réactifs auxiliaires et plasmas de contrôle Non fournis avec la trousse, ils doivent faire l’objet d’une commande séparée. Amérique et Pacifique Europe Réf. Réf. Plasma de calibrage 0020003700 0020003700 Contrôle normal 0020003120 0020003110 Contrôle anormal bas 0020003220 0020003210 Contrôle anormal haut 0020003320 0020003310 Diluant facteur 0009757600 0009757600 Contrôle de qualité Les contrôles normaux et anormaux sont recommandés pour un programme de contrôle de qualité complet.11 Les plasmas de Contrôle normal, Contrôle anormal bas, Contrôle anormal haut et de Contrôle Protéine S de la trousse sont spécifiques à ce programme. Chaque laboratoire doit établir sa propre moyenne et déviation standard et son programme de contrôle de qualité pour vérifier l'état de fonctionnement du système analytique. Les contrôles doivent être passés au minimum toutes les 8 heures en concordance avec une bonne pratique de laboratoire. Se référer au mode d’emploi de l’instrument pour des informations complémentaires. Se reporter à Westgard et al pour l’identification et la résolution des contrôles hors limites.12 Préparation Détermination des valeurs des calibrateurs et contrôles Protein S reagent: Dissoudre le contenu de chaque flacon avec 3 ml d’eau déminéralisée de type CLRW selon les normes CLSI (anciennement NCCLS) ou équivalent.9 Replacer le capuchon et agiter doucement. Assurez-vous de la complète reconstitution du produit. Conserver le réactif à 15-25°C pendant 30 minutes et mélanger par inversion avant utilisation. Ne pas agiter violemment. Protein S deficient plasma: Dissoudre le contenu de chaque flacon avec 1 ml d’eau déminéralisée de ttype CLRW selon les normes CLSI ou équivalent.9 Replacer le capuchon et agiter doucement. Assurez-vous de la complète reconstitution du produit. Conserver le réactif à 15-25°C pendant 30 minutes et mélanger par inversion avant utilisation. Ne pas agiter violemment. Protein S control plasma: Dissoudre le contenu de chaque flacon avec 1 ml d’eau déminéralisée de type CLRW selon les normes CLSI ou équivalent.9 Replacer le capuchon et agiter doucement. Assurez-vous de la complète reconstitution du produit. Conserver le réactif à 15-25°C pendant 30 minutes et mélanger par inversion avant utilisation. Ne pas agiter violemment. Les valeurs reportées dans les notices sont déterminées à partir d’essais multiples réalisés sur les systèmes de coagulation IL, en utilisant des lots spécifiques de réactif IL, et ce, par rapport à un Standard interne (Etalon Secondaire). Ce dernier est étalonné en regard du standard international Protéine S actuel, et validé à l’aide de la table de valeurs limites usuelles correspondantes. Résultats Les résultats de ProS sont reportés en secondes, en activité (%). Se référer au mode d’emploi de l’instrument pour des informations complémentaires. Conservation et stabilité des réactifs Limites de la méthode et substances interférant avec celle-ci Les résultats de ProS sur les analyseurs ACL & ACL Futura / ACL Advance ne sont pas affectés par l’héparine (Héparine Non Fractionnée ou Héparine de Bas Poids Moléculaire) jusqu’à 2 U/ml, par la bilirubine jusqu’à 15 mg/dl (150 mg/l), par l’hémoglobine jusqu’à 200 mg/dl et les triglycérides jusqu’à 1000 mg/dl (10 g/l). Les résultats de ProS sur les analyseurs ACL TOP ne sont pas affectés par l’héparine Non Fractionnée jusqu’à 1,6 U/ml ou les héparines de Bas Poids Moléculaire jusqu’à 1,8 U/ml, par la bilirubine jusqu’à 15 mg/dl (150 mg/l), par l’hémoglobine jusqu’à 200 mg/dl et les triglycérides jusqu’à 1000 mg/dl (10 g/l). Les échantillons par trop hémolysés ou lipémiques ne doivent pas être dosés. Les résultats de ProS (diminution de l’activité PS) peuvent être affectés par la présence du facteur VIIa13 (F VIIa) et par la mutation du facteur V Leiden (RPCA), chez tous les patients homozygotes.14 Dans le but d’éviter une éventuelle influence due à la présence d’une RPCA sur l’activité en ProS, les spécimens avec un taux en ProS en dehors des valeurs normales pourront être dilués manuellement au 1/2 avec le plasma déficient en Protéine S et redosés. Les résultats ainsi obtenus doivent être multipliés par le facteur de dilution. Valeurs attendues Les valeurs normales de cette étude ont été obtenues en utilisant la trousse ProS sur les appareils de coagulation IL. Appareils N % d’activité en PS ACL 78 63 - 135 ACL Futura/ACL Advance 80 67 - 132 ACL TOP Familie 49 76 - 135 * Note : l’âge et le statut hormonal peuvent affecter les valeurs normales chez la femme.15 Les normales ont été calculées comme il est recommandé par le document C28-A selon les normes CLSI.16 Ces résultats sont obtenus en utilisant un lot spécifique de réactif. Du fait qu’un grand nombre de variables peuvent affecter les résultats, nous recommandons à chaque laboratoire d’établir ses propres valeurs normales pour la ProS. Caractéristiques et performances Précision La précision intra-séries et inter-séries a été évaluée au cours d’essais multiples. ACL Moyenne (% PS) CV% (Intra-séries) CV% (Inter-séries) Contrôle Normal 91,1 3,2 7,0 Contrôle protéine S 52,5 8,2 10,2 ACL Futura/ACL Advance Moyenne (% PS) CV% (Intra-séries) CV% (Inter-séries) Contrôle Normal 90,5 3,9 5,1 Contrôle protéine S 49,1 6,6 7,6 ACL TOP Moyenne (% PS) CV% (Intra-séries) CV% (Inter-séries) Contrôle Normal 95,6 3,1 5,6 Contrôle protéine S 30,3 3,1 11,2 Corrélation: Appareils pente ordonnée à l’origine r Méthode de référence ACL 1,01 -5,883 0,982 IL Test Protéine S sur l’ACL ACL Futura 1,02 -4,890 0,984 IL Test Protéine S sur l’ACL ACL Advance ACL TOP 1,061 -4,832 0,985 IL Test Protéine S sur l’ACL Futura/ACL Advance La précision et la corrélation des résultats ont été obtenus en utilisant des lots spécifiques de réactifs et de contrôles. Linéarité : Appareils Analyseurs de coagulation IL 10 - 150 (% PS activité) Conservés à 2-8°C, les réactifs sont stables jusqu’à la date de péremption indiquée sur le flacon. Protein S reagent - La stabilité du réactif reconstitué est de: Appareils ACL® (100-7000): 2 jours à 2-8°C, 15 jours à -20°C dans le flacon d’origine ou 2 jours à 15°C sur l’analyseur. Aucune agitation mécanique n’est nécessaire. Appareils ACL 8000/9000/10000: 2 jours à 2-8°C, 15 jours à -20°C dans le flacon d’origine ou 4 heures à 15°C sur l’analyseur. Aucune agitation mécanique n’est nécessaire. Appareils ACL Futura® et ACL Advance: 12 heures à 2-8°C, 7 jours à -20°C dans le flacon d’origine ou 1 heure à 15°C sur l’analyseur. Aucune agitation mécanique n’est nécessaire. ACL TOP® : 1 heure à 15°C sur l’analyseur. Aucune agitation mécanique n’est nécessaire. Protein S deficient plasma - La stabilité du réactif reconstitué est de 4 heures à 15-25°C dans le flacon d’origine ou 7 jours à -20°C. Protein S control plasma - La stabilité du réactif reconstitué est de 4 heures à 1525°C dans le flacon d’origine ou 7 jours à -20°C. ProS - 0020002800 ITALIANO - Revisione dell’inserto 11/2009 Utilizzo Test coagulativo funzionale per la determinazione quantitativa della Proteina S libera nel plasma umano sui sistemi di coagulazione IL. Principio del metodo La Proteina S, una proteina vitamina K dipendente, è il cofattore necessario all’azione anticoagulante e profibrinolitica della Proteina C attivata.1 La Proteina S è presente nel plasma in due diverse forme: Proteina S libera (40%) e Proteina S legata al C4bBP (Complement C4b - Binding Protein) (60%). Solo la Proteina S libera è funzionalmente attiva.2 La carenza di Proteina S può essere ereditaria o acquisita. Carenze acquisite possono essere riscontrate durante la gravidanza, la terapia anticoagulante orale, l’uso di contraccettivi, nelle epatopatie, nei neonati o in altre condizioni cliniche.3,4,5 La carenza di Proteina S è generalmente associata ad un alto rischio di eventi tromboembolici venosi, specialmente in soggetti giovani. 6,7 Il kit ProS™ determina l’attività funzionale della Proteina S libera eseguendo un tempo di protrombina (PT) modificato in presenza di fattore tissutale, fosfolipidi, ioni calcio e Proteina C attivata. L’attività della Proteina S è proporzionale all’allungamento del tempo di coagulazione del Plasma carente di Proteina S a cui è stato addizionato il plasma diluito del campione in esame. Composizione Il kit ProS è composto da: R Protein S reagent (Nr. Cat. 0020002810): 3 flaconi da 3 mL di una preparazione liofilizzata contenente fattore tissutale ricombinante di coniglio, fosfolipidi sintetici, ioni calcio, Proteina C attivata, polibrene, tampone, stabilizzanti e conservanti. D Protein S deficient plasma (Nr. Cat. 0020002820): 6 flaconi da 1 mL di plasma umano liofilizzato depletato artificialmente di Proteina S. P Protein S control plasma (Nr. Cat. 0020002830): 2 flaconi da 1 mL di plasma umano liofilizzato con livelli ridotti di Proteina S. ATTENZIONE : POSSIBILE RISCHIO DI INFEZIONE Questo prodotto contiene materiale di cui è stata verificata all’origine, con metodi raccomandati dalla FDA, l’assenza dell’antigene di superficie dell’epatite B (HBsAg) e degli anticorpi anti HCV e anti HIV. Trattare come potenzialmente infetto.8 Evitare il contatto con gli occhi e con la pelle (S 24/25). Non gettare i residui nelle fognature (S 29). Usare indumenti protettivi adatti (S 36). Per l’impiego diagnostico in vitro. Preparazione Protein S reagent: Sciogliere il contenuto del flacone con 3 mL di acqua di Tipo CLRW secondo il documento CLSI (ex NCCLS) o equivalente.9 Chiudere con il tappo e miscelare delicatamente fino alla completa ricostituzione del materiale. Mantenere il reagente a 15-25°C per 30 minuti e miscelarlo nuovamente prima dell’uso. Non agitare. Protein S deficient plasma: Sciogliere il contenuto di ciascun flacone con 1 mL di acqua di Tipo CLRW secondo il documento CLSI o equivalente.9 Chiudere con il tappo e miscelare delicatamente fino alla completa ricostituzione del materiale. Mantenere il reagente a 1525°C per 30 minuti e miscelarlo nuovamente prima dell’uso. Non agitare. Protein S control plasma: Sciogliere il contenuto del flacone con 1 mL di acqua di Tipo CLRW secondo il documento CLSI o equivalente.9 Chiudere con il tappo e miscelare delicatamente fino alla completa ricostituzione del materiale. Mantenere il reagente a 1525°C per 30 minuti e miscelarlo nuovamente prima dell’uso. Non agitare. Protein S deficient plasma - Stabilità dopo la ricostituzione: 4 ore a 15-25°C nel flacone originale o 7 giorni a -20°C. Protein S control plasma - Stabilità dopo la ricostituzione: 4 ore a 15-25°C nel flacone originale o 7 giorni a -20°C. Nota: - I reagenti e i plasmi congelati devono essere scongelati a 37°C e miscelati delicatamente prima dell’uso. I componenti del kit possono essere divisi in aliquote e congelati usando provette di plastica provviste di tappo. Non ricongelare. - Al termine dei cicli lavorativi si consiglia di conservare il reagente a 2-8°C nel flacone originale per una migliore stabilità. - La condizione di conservazione dei reagenti utilizzati per la calibrazione dello strumento (non congelati o scongelati) dovrebbe essere la stessa dei reagenti utilizzati nell’analisi dei campioni e dei controlli. - Su sistemi ACL TOP è necessario eseguire la calibrazione ad ogni sessione analitica ed ad ogni gruppo di campioni. Metodo Per le metodiche e le procedure di lavoro consultare il Manuale dell’Operatore e/o il Manuale Applicativo. Preparazione dei campioni Aggiungere a nove parti di sangue venoso fresco una parte di trisodio citrato. Per informazioni aggiuntive sulla preparazione, il trattamento e la conservazione dei campioni fare riferimento al documento CLSI H21-A5.10 Reagenti ausiliari e plasmi di controllo I seguenti reattivi non sono forniti insieme al kit e devono essere ordinati separatamente. Nr. Cat. Plasma calibrante 0020003700 Controllo normale 0020003110 Controllo patologico (livello basso) 0020003210 Controllo patologico (livello alto) 0020003310 Diluente fattori 0009757600 Controllo di qualità Per un completo programma di controllo di qualità, IL raccomanda l’uso di plasmi normali e patologici.11 Il Controllo normale e i Plasmi di controllo patologico (livello basso) e (livello alto) e il Plasma di controllo proteina S sono indicati per questo programma. Ogni laboratorio dovrà stabilire i propri valori medi e le deviazioni standard per ogni controllo e stabilire le procedure di analisi. I controlli dovranno essere analizzati almeno una volta ogni 8 ore in accordo con le esigenze del laboratorio. Consultare il Manuale dell’Operatore dello strumento per ulteriori informazioni. Per l’identificazione e la risoluzione di situazioni particolari, fare riferimento all’articolo di Westgard et al.12 Calibratori e materiali di controllo I valori indicati sono stati assegnati eseguendo repliche successive sui sistemi di coagulazione IL con l’impiego di un lotto specifico di reagente verso uno Standard Interno (Aziendale) di Plasma calibrante che è riconducibile all’attuale Standard Internazionale di Proteina S, identificabile nella tabella degli intervalli di accettabilità. Risultati I valori di ProS dei campioni possono essere riportati in secondi e attività (%). Riferirsi al Manuale dell’Operatore per ulteriori informazioni. Limitazioni/sostanze interferenti I risultati del test ProS su ACL Classici e su sistemi ACL Futura/ACL Advance non sono influenzati in presenza di livelli di eparina (eparina non frazionata e a basso peso molecolare) fino a 2 U/mL, bilirubina fino a 15 mg/dL, emoglobina fino a 200 mg/dL e trigliceridi fino a 1000 mg/dL.I risultati del test ProS su sistemi ACL TOP non sono non sono influenzati in presenza di livelli di eparina (UF) fino a 1,6 U/mL, eparina (LMW) fino a 1,8 U/mL, bilirubina fino a 15 mg/dL, emoglobina fino a 200 mg/dL, trigliceridi fino a 1000 mg/dL. Nel caso di campioni emolizzati o torbidi si consiglia di ripetere il prelievo. I risultati del test ProS possono essere influenzati (riduzione dei valori di attività PS) in presenza sia di Fattore VII attivato (FVIIa)13 che della mutazione del Fattore V Leiden (APC-R)14, soprattutto in pazienti omozigoti per tale difetto genetico. Allo scopo di evitare la possibile influenza della APC-R sulla attività ProS, i plasmi dei pazienti con risultati fuori dall’intervallo di normalità dovrebbero essere diluiti 1:2 con plasma carente di proteina S e rianalizzati. I risultati ottenuti devono essere moltiplicati per il relativo fattore di diluizione. Valori attesi Utilizzando il kit ProS è stato determinato il seguente intervallo di riferimento sui sistemi di coagulazione IL. Sistema N Attività % PS ACL 100-10000 78 63 - 135 ACL Futura/ACL Advance 80 67 - 132 ACL TOP 49 76 - 135 * Nota: L’età e lo stato ormonale possono modificare l’intervallo di normalità.15 Gli intervalli di riferimento sono stati calcolati seguendo le raccomandazioni del documento C28-A del CLSI.16 Tali risultati sono stati ottenuti utilizzando un lotto specifico di reagente. A causa delle numerose variabili che possono influenzare i risultati, ogni laboratorio dovrebbe stabilire il proprio range di normalità. Prestazioni Precisione La precisione nella serie e tra ACL 100-10000 Controllo normale Controllo proteina S ACL Futura/ACL Advance Controllo normale Controllo proteina S ACL TOP Controllo normale Controllo proteina S Correlazione: Sistema pendenza ACL 100-10000 1,01 ACL Futura 1,02 ACL Advance ACL TOP 1,061 serie è stata verificata eseguendo numerose repliche. Media (% PS) CV% (nella serie) CV% (Totale) 91,1 3,2 7,0 52,5 8,2 10,2 Media (% PS) CV% (nella serie) CV% (Totale) 90,5 3,9 5,1 49,1 6,6 7,6 Media (% PS) CV% (nella serie) CV% (Totale) 95,6 3,1 5,6 30,3 3,1 11,2 intercetta -5,883 -4,890 r Metodo di riferimento 0,982 Proteina S coagulativa IL su ACL 0,984 Proteina S coagulativa IL su ACL -4,832 0,985 Proteina S coagulativa IL su ACL Futura/ACL Advance I risultati dei test di precisione e di correlazione sono stati ottenuti usando uno specifico lotto di reagenti e controlli. Linearità: Sistema Sistemi di coagulazione IL 10 - 150 (Attività % PS) Conservazione e stabilità dei reagenti I reagenti sigillati sono stabili a 2-8°C fino alla data di scadenza stampata sull’etichetta. Protein S reagent - Stabilità dopo la ricostituzione: Strumenti ACL 100-7000: 2 giorni a 2-8°C, 15 giorni a -20°C nel flacone originale o 2 giorni a 15°C sullo strumento. Non è richiesto l’agitatore per la miscelazione. Strumenti ACL 8000/9000/10000: 2 giorni a 2-8°C, 15 giorni a -20°C nel flacone originale o 4 ore a 15°C sullo strumento. Non è richiesto l’agitatore per la miscelazione. Strumenti ACL Futura e ACL Advance: 12 ore a 2-8°C, 7 giorni a -20°C nel flacone originale o 1 ora a 15°C sullo strumento. Non è richiesto l’agitatore per la miscelazione. Sistemi ACL TOP: 1 ora a 15°C sullo strumento. Non è richiesto l’agitatore per la miscelazione. ProS - 0020002800 PORTUGUÊS - Revisão do folheto 11/2009 Aplicação Técnica de coagulação funcional automatizada para a determinação quantitativa da Proteína S livre em plasma humano citratado nos Sistemas de Coagulação da IL. Princípio A Proteína S é uma glicoproteína vitamina K-dependente, que actua como cofactor da Proteína C Activada, aumentando os seus efeitos anti-coagulantes e profibrinolíticos.1 A Proteína S está presente no plasma sob duas formas: Proteína S livre (40%) e Proteína S ligada à proteína transportadora da fracção C4b do complemento (60%). As duas formas estão em equilíbrio dinâmico e apenas a Proteína S livre tem actividade biológica.2 Os défices de Proteína S podem ser congénitos ou adquiridos. Estes últimos podem ser observados: durante a gravidez, em terapia anti-coagulante oral (TAO), uso de anti-contraceptivos orais, hepatopatías, recém nascidos, assim como em outras situações clínicas.3,4,5 Os défices de Proteína S estão associados a um risco maior de tromboembolismo venoso, especialmente em pessoas jovens.6,7 O ensaio ProS™ determina a actividade funcional da Proteína S livre medindo o grau de prolongamento do Tempo de Protrombina (TP) na presença de factor tecidular, fosfolípidos, iões de cálcio e Proteína C activada. A actividade da Proteína S é proporcional ao prolongamento do tempo de coagulação do Plasma Deficiente em Proteína S ao qual se incluiu a amostra de paciente diluída. Composição O kit de ProS é composto por: R Protein S reagent (Núm. Cat. 0020002810): recipientes 3 x 3 mL de uma preparação liofilizada que contém factor tecidular recombinante de coelho, fosfolípidos sintéticos, iões de cálcio, Proteína C activada, polibrene, tampão, estabilizadores e conservantes. D Protein S deficient plasma (Núm. Cat. 0020002820): recipientes 6 x 1 mL de plasma humano liofilizado, do qual se eliminou artificialmente a Proteína S. P Protein S control plasma (Núm. Cat. 0020002830): recipientes 2 x 1 mL de plasma humano liofilizado que contém uma concentração baixa de Proteína S PRECAUÇÕES E ADVERTÊNCIAS: O material utilizado neste produto foi analisado com testes aprovados pela FDA e verificou-se a ausência de reacção ao Antigénio de Superfície da Hepatite B (HBsAg), aos anticorpos anti-HCV e anti-HIV 1/2. No entanto, deve-se manipular com precaução, como potencialmente infeccioso.8 Evitar o contacto com a pele e os olhos (S 24/25). Não deitar os resíduos no esgoto (S 29). Usar equipamento de protecção adequado (S 36). Este reagente destina-se a utilização em diagnóstico in vitro . Preparação Protein S reagent: Dissolver o conteúdo de cada recipiente com 3 mL de água tipo CLRW da CLSI (anteriormente NCCLS) ou equivalente.9 Fechar o recipiente e homogeneizar suavemente. Verificar se o produto fica completamente dissolvido. Conservar o reagente entre 15 e 25°C durante 30 minutos. Misturar invertendo o recipiente, antes de utilizar. Não agitar. Protein S deficient plasma: Dissolver o conteúdo de cada recipiente com 1 mL de água tipo CLRW da CLSI ou equivalente.9 Fechar o recipiente e homogeneizar suavemente. Verificar se o produto fica completamente dissolvido. Conservar o reagente entre 15 e 25°C durante 30 minutos. Misturar invertendo o recipiente, antes de utilizar. Não agitar. Protein S control plasma: Dissolver o conteúdo de cada recipiente com 1 mL de água tipo CLRW da CLSI ou equivalente.9 Fechar o recipiente e homogeneizar suavemente. Verificar se o produto fica completamente dissolvido. Conservar o reagente entre 15 e 25°C durante 30 minutos. Misturar invertendo o recipiente, antes de utilizar. Não agitar. Notas: - Os plasmas e os reagentes congelados deverão ser descongelados a 37°C. Homogeneizar suavemente antes de utilizar. Antes de congelados, os produtos deverão ser repartidos em alíquotas, usando tubos de plástico com tampa. Não recongelar. - Para optimizar a estabilidade, devem retirar-se os reagentes dos equipamentos e devem conservar-se a 2-8°C nos recipientes originais bem tapados. - As condições de armazenamento dos reagentes utilizados na preparação da curva de calibração devem ser as mesmas que, para os reagentes utilizados na análise de amostras/controlos (não congelados ou descongelados). - Na Família ACL TOP recomenda-se a calibração em cada sessão de trabalho e indicação do lote utilizado na sessão de análise. Limitações/interferências Seguir as instruções da técnica de acordo com o Manual do Operador e/ou com o Manual de Aplicações dos Instrumentos IL. Concentrações de heparina (heparina não fraccionada ou heparina de baixo peso molecular) até 2 U/mL, de bilirrubina até 15 mg/dL, de hemoglobina até 200 mg/dL e de triglicéridos até 1000 mg/dL não alteram os resultados ProS na Família ACL e ACL Futura/ACL Advance Systems. Os resultados ProS na Família ACL TOP não são afectados pela heparina não fraccionada (UF) até 1.6 U/mL, heparina (BPM) até 1.8 U/mL, bilirrubina até 15 mg/dL, hemoglobina até 200 mg/dLe triglicéridos até 1000 mg/dL. Não utilizar amostras hemolisadas e turvas. Os resultados de ProS podem ser afectados (diminuindo a actividade da PS) tanto pela presença do Factor VII activado na amostra (Factor VIIa)13, como pela mutação do Factor V (Factor V Leiden/APC-R)14, sobretudo em pacientes homozigóticos para esta mutação genética. Para evitar uma possível influência da APC-R nos valores da actividade da ProS obtidos, as amostras com resultados fora do limite normal devem ser diluídas manualmente de 1:2 com Plasma deficiente em Proteína S e devem ser novamente analisadas. Os resultados impressos devem ser multiplicados por 2 para corrigir a diluição. Recolha e preparação de amostras Valores esperados Adicionar nove partes de sangue extraído recentemente por punção venosa a uma parte de citrato trissódico. Para efectuar a recolha, o manuseamento e a conservação da amostra devem seguir-se as recomendações referidas no documento H21-A5 da CLSI.10 Efectuou-se um estudo do intervalo de normalidade utilizando o kit da ProS nos Sistemas de Coagulação da IL. Sistema N Actividade (%) PS Familia ACL 78 63 - 135 ACL Futura/ACL Advance 80 67 - 132 Familia ACL TOP 49 76 - 135 * Nota: A idade e a situação hormonal podem afectar o intervalo normal para o sexo feminino.15 Os limites foram calculados de acordo com as recomendações referidas no documento C28-A da CLSI.16 Os resultados foram obtidos utilizando o mesmo lote de reagente. Devido às diversas variáveis que podem afectar os resultados, cada laboratório deve estabelecer o seu próprio intervalo de normalidade. Método de ensaio Reagentes adicionais e plasmas de controlo Este kit não inclui os produtos abaixo mencionados, pelo que deverão ser pedidos em separado: Américas e Pacífico Europa Núm. Cat. Núm. Cat. Plasma de Calibração 0020003700 0020003700 Controlo Normal 0020003120 0020003110 Controlo Anormal Baixo 0020003220 0020003210 Controlo Anormal Alto 0020003320 0020003310 Diluente de factores 0009757600 0009757600 Controlo de Qualidade Para realizar um programa completo de Controlo de Qualidade, recomenda-se a utilização dos controlos normais e anormais.11 Os controlos adequados para este programa são o Controlo Normal, o Controlo Anormal Baixo, o Controlo Anormal Alto e o Plasma Controlo de Proteína S. Cada laboratório deve estabelecer a sua própria média e desvio padrão e deve efectuar um programa de controlo de qualidade para monitorizar os seus resultados. Os controlos devem ser analisados uma vez em cada turno de 8 horas, de acordo com a norma vigente no Laboratório. Ver Manual do Operador para mais informações. Consultar a publicação de Westgard e col. para identificar e solucionar situações anormais do Controlo de Qualidade.12 Determinação dos valores dos calibradores e materiais de controlo Os valores referidos foram determinados através de diversas análises efectuadas nos Sistemas de Coagulação da IL, utilizando um lote específico de reagente IL e comparando com um Padrão interno da IL de acordo com os Padrões Internacionais de Proteína S actualmente utilizados, identificados na tabela do intervalo de aceitação. Resultados Os resultados de ProS são expressos em segundos e Actividade (%).Consultar o Manual do Utilizador para mais informações. Características Técnicas Precisão A precisão intra-série e inter-série foi realizada em diferentes séries. Familia ACL Média (% PS) CV% (Intra-série) CV% (Total) Controlo Normal 91,1 3,2 7,0 Controlo Proteína S 52,5 8,2 10,2 ACL Futura/ACL Advance Média (% PS) CV% (Intra-série) CV% (Total) Controlo Normal 90,5 3,9 5,1 Controlo Proteína S 49,1 6,6 7,6 Familia ACL TOP Média (% PS) CV% (Intra-série) CV% (Total) Controlo Normal 95,6 3,1 5,6 Controlo Proteína S 30,3 3,1 11,2 Correlação: Sistema pendente intersecção r Método Comparativo Familia ACL 1,01 -5,883 0,982 IL Proteína S de coagulação em ACL ACL Futura 1,02 -4,890 0,984 IL Proteína S de coagulação em ACL ACL Advance Familia ACL TOP 1,061 -4,832 0,985 IL Proteína S de coagulação em ACL Futura/ACL Advance A precisão e os dados de correlação foram obtidos utilizando lotes específicos de reagentes e de controlos. Linearidade: Sistema Sistemas de Coagulação IL 10 - 150 (Actividade % PS) Conservação e estabilidade dos reagentes Os recipientes que não tenham sido abertos e tenham sido conservados entre 2-8°C, são estáveis até à data de validade indicada no rótulo. Protein S reagent - Estabilidade após reconstituição: - Sistemas ACL (100-7000): 2 dias a 2-8°C, 15 dias a -20°C no recipiente original ou 2 dias a 15°C no aparelho. Não é necessário agitar. - Sistemas ACL 8000/9000/10000: 2 dias a 2-8°C, 15 dias a -20°C no recipiente original ou 4 horas a 15°C no aparelho. Não é necessário agitar. - Sistemas ACL Futura e ACL Advance: 12 horas a 2-8°C, 7 dias a -20°C no recipiente original ou 1 hora a 15°C no aparelho. Não é necessário agitar. - Família ACL TOP®: 1 hora a 15ºC no aparelho. Não é necessária agitação. Protein S deficient plasma - Estabilidade após reconstituição: 4 horas a 15-25°C no recipiente original ou 7 dias a -20°C. Protein S control plasma - Estabilidade após reconstituição: 4 horas a 15-25°C no recipiente original ou 7 dias a -20°C. Instrumentation Laboratory Company - Bedford, MA 01730-2443 (USA) Instrumentation Laboratory SpA - V.le Monza 338 - 20128 Milano (Italy) 303049 R9 11/2009 ProS Printed Insert Sheet: Revision: Issued: C.O.: 0020002800 303049 R9 November 2009 412182 LANGUAGES ENGLISH DEUTSCH ESPAÑOL FRANÇAIS ITALIANO PORTUGÛES TECHNICAL SPEC'S PAPER: SIZE: PRINT: PRINT COLOR: White paper, 50-60 g/m2 weight. 11 x 24" (280 x 609 mm.). Front/Back. Front - Top band Green Pantone 382, all remaining type in black. Back - All type in black. Instrumentation Laboratory Company - Bedford, MA 01730-2443 (USA) Instrumentation Laboratory SpA - V.le Monza 338 - 20128 Milano (Italy) 303049 R9 11/2009