Fiche de vie glider potame (Slocum #070)

Transcription

Fiche de vie glider potame (Slocum #070)
Profil de vie Glider Slocum#070
12/11/2008 17:37
Fiche de vie glider potame (Slocum #070)
Date
12/01/07
Lieu
LPO Brest
Objet
Shipped by WRC via Nantes
22-23/01/07
LPO, Brest
Power on the glider for the first time
28/01/07
LPO, Brest
30/01/07
IMEDEA, Calanova
Ship to Mallorca, Imedea for Med
deployment
Prepare and ballast
31/01/07
IMEDEA, Calanova
01/02/07
Off Soller, Mallorca
02/02/07
04/02/07
18/02/07
02/03/07
Document de référence
Observations
Airshipped from Boston / Paris / Nantes and then
truck Nantes to Brest. Nantes is the closest airport
with cargo facilities. Due to the box sizes, it can only
go in a cargo. Open the box and visually inspect the
glider and the box content. Nothing to note.
Compte rendu de préparation et Activate the SIM Iridium card. Change the autoexec
déploiement d’un planeur Spray with glider name and phone number to call. Check
functions : pump, GPS, Iridium calls. Check
et du Slocum profond du LPO
dockserver
en Méditerranée Occidentale
dans le cadre de MERSEA
WP3, T.Terre, P.Lherminier,
February 2007
idem
Door to door shipping by special truck
idem
Check CTD + oxygen sensor
Check ballast for potame
idem
idem
Off Soller, Mallorca
First try to pdeploy the glider but cancelled
due to bad weather
Deploy the glider but unsuccesfull. Decide
to get it back to Brest for pressure tank
tests.
Verification of pool water characterisitics showed a
significance diffrence with the previously assumed
numbers.
Deepy alone was deployed
IMEDEA, Calanova
LPO, Brest
LPO, Brest
Prepare for shipping back to Brest
Glider back to lab
Open and separate the 3 sections
idem
During the first try of the mission at 30 m, the glider
never came back to the surface due to a bad
functionning of the vehicle pressure sensor. Vaseline
was introduced in the capillar for corrosion problem
but it prevents the pressure to be correctly transmitted
to the sensor. Back to shore with the glider.
idem
Door to door truck shipping with the same carrier
This is to change the vehicule pressure sensor capillar
tube
1/5
Profil de vie Glider Slocum#070
12/11/2008 17:37
05/03/07
LPO, Brest
Dismount tube and check the pressure
sensor
12/03/07
LPO, Brest
13-16/03/07
LPO, Brest
19-20/03/07
LPO, Brest
Replace the capillar tube and check the
overall pressure sensor circuit. Close the
glider and put a vacuum at 7.4 inHg
Check the vacuum is not decreasing .
record Vacuum and internal temperature.
Open the glider again to prepare the
pressure tests by setting a cable for
communication thru the pressure tank
C. Jones visit to LPO
Report on Clayton Jones visit to
Pressure tank tests while running the pump Ifremer/LPO, T.Terre, march
at 250 db and 1000 db
2007
Pool trials
12-20/04/07
OOB, Banyuls
Preparation and deployment with N/O
Nereis
17/04-18/06
Western Mediterranean
Contribution to the convection monitoring
with NOC and LOCEAN teams
Compte-rendu de récupération
d’un planeur Spray et de
préparation et déploiement du
Slocum profond du LPO en
Méditerranée Occidentale dans
le cadre de Mersea WP3,
T.Terre, mai 2007
Report on operations with
Slocum Potane (unit #70) in
Western Mediterranean Sea.
T.Terre, July 2007
2/5
Vaseline clearly seen at the entrance of the pressure
sensor. Check pressure variation with vacuum pump
and nitrogen injection (adtest)
Same vacuum pump and nitrogen use
Vehicle pressure sensor OK
Pump ran OK in presure tank
Pool tests showed offset between vehicle pressure
sensor and SBE CTD pressure sensor
Pump was set OUT OF SERVICE due to an OUT
OF DEADBAND error (later corrected by soft)
Shipping door to door with Ifremer van
Start in mission without DO sensor running. Set up
Later by changing sbdlist.dat
Diificulties to connect with Iridium : got an abort
MS_ABORT_DEVICE_ERROR 2007-0506T18:43:57
18/05 Leak detect Abort; Recovery and check; Loose
screws shortening the leak detect sensor !
Redeploy on 19/04
Start in mission without DO sensor running. Set up
later by changing sbdlist.dat
Difficulties to connect with Iridium : got an abort
06/05 : MS_ABORT_DEVICE_ERROR 2007-0506T18:43:57
20/05 : Deep engine out of service
OOD:oil_volume:OUT OF DEADBAND:
M_su:225.4825 C_su:266.3760
change
f_de_oil_vol_deadz_width from 40 to something
bigger
10/06 : abort due to a bad syntwx in a mafile.
Profil de vie Glider Slocum#070
12/11/2008 17:37
18/06/07
OOB, Banyuls
Recovery
20-21/09/07
LPO, Brest
Pressure tank tests and batteries discharge
20/10-4/11
2nd EGO Workshop and Potame send as instrument for the school
Glider School, Mallorca
Shipping to USA
Potame is sent back to WRC
Delivered to WRC
Thanksgiving around !
Shipping WRC to LPO
Delivered to LPO
15-16/11/07
26/11/07
21/12/07
15/01/08
22-23/01/08
02/02/08
Shipping Brest-Toulon
During all the deployment low success in Iridium
comms ~20 %
low H-moment as first cause 2nd
cause could be turned air bladder ?
Questions raised :
Compte rendu de récupération
1. Test of Iridum link
d’un planeur Slocum profond du
2. purge of deep engine
return to WRC ?
LPO en Méditerranée
3. air bladder attitude, does oxygen sensor
Occidentale dans le cadre de
have an effect ?
MERSEA WP3, T.Terre, juin
4. Cycles the pump at 1000 db in pressure tank
2007
to empty the lithium batteries
5. Change pressure hull from aluminium to
composite ?
Report on Iridium, pump test
Run the pump for 140 cycles at 100bar with the
and batteries discharge after
batteries used during the experiment.
Medterranean Sea dployment
Discharge the batteries on resistive loads
Transport Brest-Mallorc-Brest by Ifremer van, driver
TT.
Air purge + switch to composite hull
Compte-rendu de préparation et
déploiement d’un planeur
Slocum profond du LPO en
Méditerranée occidentale dans
le cadre de MERSEA WP3.,
T.Terre, février 2008
3/5
List of upgrades :
- composite hulls for FW/AFT sections
- peek for CTD pressure sensor circuit
- new CTD pressure transducer block
- new vacuum plug
- new vacuum tool
- suppress pinger compense weight by lead bottle
- revise altimeter (170 kHz) block
- change some components
- new sofware version
- full tests in Ashumet Pond
Shipping Brest-Toulon on 22-23/01 by Ifremer van
Banyuls on 26/01
Profil de vie Glider Slocum#070
12/11/2008 17:37
23-25/01/08
Ifremer, Toulon
26/01/08
Shipping Toulon Banyuls
OOB, Banyuls
End of preparation for deployment
29/01/08
Off shore Banyuls
Deployment with Nereis II
idem
30/0101/02/08
29/01-07/04
OOB, Banyuls
Pilot the glider while PLH is away
idem
Western Mediterranean
Pilots PLH & TT
27-28/03/08
Shipping Brest-Banyuls
28/03/08
OOB, Banuyls
27/01/0828/01/08
Preparation for EGO deployment in gulf of
Lion
Training on air purge in the field
idem
idem
idem
Shipping Toulon-Banyuls on 26/01 by Ifremer van
2 wires on connector 5 points i.e. cable for the optode
are broken. No tool to redo the connector
soldier
the wires ?!
No problem except a false alarm with the altimeter.
1 CTD with 5 samples at 5 depths
Dives to 300, 500, 800 and 1000 m.
Problems encountered
- 04/02 abort device_error due to optode ; turn it
off
- 16/02 abort same_depth : landing on the bottom
during the slope approach off Barcelona
- 27/02 abort device_error : link to the pump but
happens once and did nothing
- 04/03 abort device_error : link to the pump send
a command to reset a flag on Potame
!put x_de_avg_oil_vol_err_on_descent 0
- 05-06-07/03 abort no_tickle : Potame suddenly
too light cant dive; very bad waetaher, not able
to connect for 4 days
- 08/04 abort overdepth : still not able to dive
change the pump volume to fix the case
Compte-rendu de récupération
de 2 planeurs Slocum profonds
en Méditerranée occidentale
dans le cadre de MERSEA
WP3., T.Terre, février 2008
Recovery of Himilcon with Nereis II
idem
4/5
Recovery of Potame and Himilcon
Problems with Himilcon
- its 2 wings were broken
- Heavy corrosion !!
Profil de vie Glider Slocum#070
12/11/2008 17:37
07/04/08
OOB, Banyuls
08/04/08
18/04/08
Shipping Banyuls –
Brest
LPO, Brest
10/11/08
Shipping to USA
After a week, waiting for the good weather
window, recovery of Potame with Nereis II
idem
idem
Pool test : deploy Potame together with an Rapport de Stage J.Burdallet.
SBE25 to post calibrate the CTD. Oxygene
and salinity samples.
Need repair on :
- Air purge
- Optode connexion
- Big connector
5/5
Problem with Potame
- Its 2 wings were broken I is the origin of the
problem encountered 05-08/03/08 during the
bad weather.
- Corrosion on the nose
Ifremer van
Bend a pin on the big connector !