Thermo Scientific Biological Safety Cabinets
Transcription
Thermo Scientific Biological Safety Cabinets
Tec h ni c a l N ote: TN B S C TCO N A Thermo Scientific Biological Safety Cabinets Managing a Facility’s Total Cost of Operation Key Words Total Cost of Operation (TCO), Biological Safety Cabinets (BSCs), Hoods, Fume Hoods, Energy Efficiency Goal Managing groups of biological safety cabinets (BSCs) in a biomedical research, healthcare or academic facility requires three strategies to optimize the total cost of operation. These strategies address both performance and cost. An inexpensive BSC that does not provide the needed cleanliness and containment has no value; conversely, a BSC that provides unneeded or unused protection is a waste of important resources that could be better utilized elsewhere. Strategy One: The right protection for your application Biological safety cabinets (BSCs), clean benches, chemical fume hoods and additional airflow cabinetry are commonly called “hoods.” Although laboratory hoods are often lumped together, each provides very different types of protection with different cost structures. construction and do not have internal fans or filters. The major operating cost associated with fume hoods is heating, cooling and conditioning the air required to replace the expelled laboratory air. A Mills and Sartor1 study on fume hoods estimated the annual cost of external exhaust at $4.50 per cfm per year. A typical fume hood with an opening 3 ft wide and 18 inches tall requires approximately 450 cfm and costs over $2,000 per year in energy. If biological containment or protection is required, then HEPA-filtered, non-exhausted hoods or Type A2 BSCs are necessary. If the containment or protection required is from volatile chemicals or gas, then externally-exhausted chemical hoods or Type B2 BSCs are necessary. Figure 1 illustrates the level of protection required for the various types of airflow cabinets. From a cost standpoint, the decision that a BSC must provide protection from gases and volatile chemicals, in addition to biological hazards, increases the overall cost of the BSC significantly. For safety and efficiency, it becomes vital to determine precisely what type of protection is needed. Airflow Cabinets & Protection Level Protection Level Airflow Cabinet Type Class II, Type A2 BSC (Non-Exhausted, HEPA-Filtered) Class II, Type B2 BSC (Externally Exhausted, HEPA-Filtered) Chemical Fume Hood (Externally Exhausted) Personal * * * Environment * * * * * * * Open Air Lab Bench Product Laminar Flow Clean Bench (HEPA-Filtered) * Cross-contamination Figure 1: Select the type of airflow product required based on whether or not biohazards or toxic chemicals are used in the application. Biological safety cabinets, laminar flow clean benches and cleanrooms capture and control “particles” including biological agents through HEPA filtration. With proper maintenance, most of these filtration-based devices can last five to ten years between filter changes. Chemical fume hoods use airflow to capture, dilute and expel hazardous gases and fumes from the laboratory or work area. Chemical fume hoods tend to be simpler in * Based on internal data, 2011. Data on file. Total Cost of Operation over 15 years $ 20,000 $ 15,000 $ 10,000 $ 5,000 $ Non-Exhausted Externally BSC Exhausted BSC Externally Exhausted Chemical Fume Hood Figure 2: Estimated equipment costs for energy and maintenance over 15 years. Externally exhausted models double costs.* Figure 2 provides a rough comparison of total costs in energy and upkeep for the three types of containment equipment discussed: Non-Exhausted BSCs, Externally Exhausted BSCs, Externally Exhausted Chemical Fume Hood. Note how the decision to provide protection from both biological hazards and volatile toxic chemicals, which requires the equipment to be externally exhausted, more than doubles the total cost of operating the BSC. Strategy Two: Assess your usage needs With the exception of external exhaust, the major ongoing costs of a BSC are associated with usage. When the BSC is operating; the fans are spinning, the fluorescent lights are on, the filters are loading up with captured particles and motors are wearing down. None of this happens when the BSC is completely turned off. Two BSC Usage Groups Weekly Hours BSC in Operation 200 150 100 50 0 10 203040 5060 Weekly Hours Used Figure 3: Results showing hours/week when users are working in BSCs (yellow) and hours BSC blowers were left operating regardless of active usage (blue). Average weekly use is 18 hours/week.* Yellow = Only on when used Blue = 24/7 ‘On’ or “Power User” Some BSCs are used only occasionally when the particular application requires the cleanliness and containment of a HEPA-filtered BSC. Other BSCs are constantly in use because of laboratory demands. Figure 3 shows the results of a BSC survey undertaken by Thermo Fisher Scientific to better understand BSC usage. The horizontal x-axis represents the hours per week the users reported physically working inside the BSC. The vertical y-axis represents the hours per week the users reported the BSC blowers were left in operating mode. Notice that BSC usage falls into two distinct categories. The first category (yellow) are units where the BSC blowers are only on when the BSC is in use (i.e., the BSC is turned off when it is non-operational). The second category (blue) are units where the BSC blowers are on * Based on external field study, 2011. Data on file. essentially 24/7, regardless of use. In surveys to date, we have found 27 to 30% of the BSCs fall into this blue category, which we call “Power Users.” Regardless of the number of hours the BSCs were in operation, all users reported < 60 hours of use per week. The average use was < 18 hours. A part of an overall strategy to maximize the cleanliness and containment provided by a facility’s BSCs while optimizing the costs requires recognition that the usage and costs can vary significantly. Some energy efficient BSCs available on the market today operate at 75% lower energy consumption levels than commonly used, traditional BSCs. Therefore, “Power users” should consider upgrading to newer BSC technology for maximum reduction in their facility’s overall operating costs. Strategy Three: Timing is everything The costs of operating a BSC are not evenly distributed over the 15 or 20 year lifespan of the unit. In the first year, electricity and cooling are the only additional costs. At some point, the filters and/or motors will need replacement. The cost for this is typically between $1,000 and $2,000 or even more. For the traditional BSC, filters and motor replacement are approximately 35% of the total cost of operation over 15 years. Estimated Annual BSC Expense for Power Users* Annual operating costs 2 $ 2,500 $ 2,000 $ 1,500 $ 1,000 $ 500 $ 123456789101112 1314 15 Number of Years in Use Figure 4: Estimated operating cost per year for “Power User” with cost being constant except for years when maintenance is required.* Figure 4 shows a typical pattern of expenditure for a BSC Power User. Note the annual expenses are the same from year to year, unless the filters needs replacing (as in years seven and fourteen) or the motor needs replacing (as in year ten). Replacing a BSC rather than spending the money to mend significant components may be the most cost effective way to transition from a costly, inefficient BSC to a newer, more cost effective BSC. BSC: Five Steps to Managing TCO These strategies can be implemented as part of an overall approach using these steps: 1. Conduct a BSC inventory survey, capturing: - unit model - location and identification - usage (hours per day, days per week) - electrical energy consumption - external exhaust volume - room exhaust volume - application (biohazards, volatile toxic chemicals, etc.) Note that the most costly non-externally exhausted BSCs How Thermo Scientific BSCs Can Reduce ITALIAN thermoscientific.com on the far right of the chart (shown in red) are all Class Your Facility’s TCO: II, Type B2 BSCs; the next most costly BSCs (shown in • Scientific Long Inc. life components – expected motor life >100,000 © 2012 Thermo Fisher Scientific Inc. Tutti i diritti riservati. Tutti gli altri marchi sono di proprietà di Thermo Fisher e delle sue filiali. Specifiche, condizioni e tariffe subire variazioni. Non tutti i prodotti sono disponibili in tuttiA2 i paesi. Per maggiori dettagli consultare il rivenditore locale motor changes over the life of the yellow) arepossono externally exhausted Class II, Type and hours means fewer B1 BSCs. This supports the value of the first strategy. cabinet Australia +61 39757 4300 India +91 22 6716 2200 Svizzera +41 44 454 12 22 Taking the nominal 4 ft width as an example, a typical Austria +43 1 801 40 0 Italia +39 02 95059 554 UK/Irlanda +44 870 609 9203 • Mini-pleat HEPA filters – more surface area equates to Belgio +32 53 73 42 41 Olanda +31 76 579 55 55 USA/Canada +1 866 984 3766 Class II, Type B2 BSC requires an external exhaust of Cina +86 21 6865 4588 or Nuova Zelanda +64 9 980 6700 longer filter life 665 II, Type A2 BSC of +86 10cfm 8419and 3588 a non-exhausted Class Europa del Nord/CIS Altre Nazioni Asiatiche +852 2885 4613 Francia +33 2 2803 2180 +358 9 329 10200 Nazioni non in elenco +49 6184 90 6000 the same size requires 360 cfm. • Balanced – downflow and inflow load evenly Germania +49 6184 90 6000 Russia +7 812 703 42 15 o +33 2filters 2803 2180 Giappone +81 45 453 9220 Spagna/Portogallo +34 93 223 09 18 eliminating premature replacement The cost of the average externally exhausted BSC was more than fourteen times greater than the average of the • Energy efficient DC motors – reduce energy non-externally exhausted BSC. External exhaust is needed consumption by 75% GERMAN thermoscientific.com to provide needed protection to BSC users also working • Night Set-Back mode – energy saving mode for low with Given Alle that external exhaust © 2012 volatile Thermo Fishertoxic Scientificchemicals. Inc. Alle Rechte vorbehalten. anderenWarenzeichen sind Eigentum von Thermo Fisher Scientific Inc. bzw. nationalen periods Niederlassungen. Änderungen von Spezifikationen, Bedingungen und Preisen vorbehalten. Nicht alle Produktesindusage in allen Ländern erhältlich. Ausführliche Informationen erhalten Sie bei Ihrer Thermo Fisher Vertretung. Annual costs Australien +61 39757 4300 Indien +91 22 6716 2200 Russland +7 812 703 42 15 Operating Cost of Energy and 554 Exhaust for Class II BSCs Surveyed+34 (Ranked from Belgien +32 53 73 42Annual 41 Italien +39 02 95059 Spanien/Portugal 93 223 09 18 low to high) China +86 21 6865 4588 or Japan +81 45 453 9220 Schweiz +41 44 454 12 22 $ 8419 6,0003588 Class II, Type A2 (Non-Exhausted) +86 10 Niederlande +31 76 579 55 55 UK/Irland +44 870 609 9203 Deutschland NationaleClass kostenlose Neuseeland +64 9 980 6700 USA/Kanada +1 866 984 3766 II, TypeHotline A2 (Externally Exhausted) $ 5,000 0800 1 536 376 Nordeuropa/Baltikum/CIS Class II, Type B2 (Externally Exhausted) Deutschland International +358 9 329 10200 Andere asiatische Länder +852 2885 4613 $ 4,000 +49 6184 90 6000 Österreich +43 1 801 40 0 Andere Länder +49 6184 90 6000 Frankreich +33 2 2803 2180 oder +33 2 2803 2180 $ 3,000 $ 2,000 300 285 270 255 240 225 210 195 180 165 150 135 120 105 90 75 FRENCH 60 45 30 thermoscientific.com 15 $ 1,000 © 2012 Thermo Fisher Scientific Inc. Tous droits réservés. Les autres marques déposées sont des marques commerciales ou déposées de Thermo Fisher Scientific Inc et de filiales. Les caractéristiques, conditionsand et tarifs sont susceptibles d’être modifiés. Tous lesof produits ne sont pas disponibles dansType tous lesA2 (blue) models are least Figure 5:ses Annual total costs (energy exhaust) for different types BSCs. Non-exhausted pays. Pour tout renseignement, veuillez vous adresser à votre distributeur local. expensive to operate while externally exhausted Type B2 for chemical protection (red) are highest. * Allemagne +49 6184 90 6000 France +33 2 2803 2180 Russie +7 812 703 42 15 Australie +61 39757 4300 Inde +91 22 6716 2200 Suisse +41 44 454 12 22 Autriche +43 1 801 40 0 Italie +39 02 95059 554 UK/Irlande +44 870 609 9203 Belgique +32 53 73 42 41 Japon +81 45 453 9220 USA/Canada +1 866 984 3766 * Based on external field study, 2011. Data on file. Chine +86 21 6865 4588 or Nouvelle-Zélande +64 9 980 6700 +86 10 8419 3588 Pays Bas +31 76 579 55 55 Autres pays d’Asie +852 2885 4613 1 Energy use and forPays laboratory fume hoods, Evan Mills,Autres DalePays Sartor, Lawrence Espagne/Portugal +34savings 93 223 09potential 18 nordiques/baltes/CEI +49 6184 90 6000 Berkeley National Laboratory, +358 9 of 329California, 10200 ou +33 2 2803 2180 Energy Analysis Department, University 2003 thermoscientific.com thermoscientific.com/bscnewlab © 2012 Thermo Inc.Inc. All rights reserved. All trademarks are theare property of ThermoofFisher Scientific and its subsidiaries. Specifications,Specifications, ThermoFisher FisherScientific Scientific All rights reserved. All trademarks the property Thermo FisherInc. Scientific Inc. and its subsidiaries. terms and and pricing pricing are NotNot all products are available in all countries. Please consult localyour saleslocal representative for details. for details. terms aresubject subjecttotochange. change. all products are available in all countries. Please your consult sales representative Australia +61 39757 4300 Austria +43 1 801 40 0 Belgium +32 53 73 42 41 China +86 21 6865 4588 or +86 10 8419 3588 France +33 2 2803 2180 Germany national toll free 0800 1 536 376 Germany international +49 6184 90 6000 TNBSCTCONA 0612 India toll free 1800 22 8374 India +91 22 6716 2200 Italy +39 02 95059 554 Japan +81 45 453 9220 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Nordic/Baltic/CIS countries +358 9 329 10200 Russia +7 812 703 42 15 Spain/Portugal +34 93 223 09 18 Switzerland +41 44 454 12 22 UK/Ireland +44 870 609 9203 USA/Canada +1 866 984 3766 Other Asian countries +852 2885 4613 Countries not listed +49 6184 90 6000 3 Tec h ni c a l N ote: TN B S C TCO N A cost is the single greatest contributor to the operational cost to this population of Class II BSCs, verifying the need and identifying the amount of external exhaust needed ©2. 2012 Thermo Fisher Scientific Reservados todos los of derechos. Las otras marcas Identify the top Inc. 20% in terms operating costcomerciales son propiedad de Thermo Fisher Scientific Inc. y sus could enmake tremendous filiales. Características, condiciones y precios están sujetos a posibles cambios. Algunos productos no están disponibles todos losapaíses. Consulte los difference to the bottom line of detalles con su representante comercial local. the organization. 3. Review application requirements with environmental health safety Alemania +49 and 6184 90 6000 Francia +33 2 2803 2180 Rusia +7 812 703 42 15 As Figure 5 illustrates, 70% of the operating cost of the Australia +61 39757 4300 Holanda +31 76 579 55 55 Suiza +41 44 454 12 22 Austria +43 1 801 40 0 India +91 22 6716 2200 UK/Irlanda +44 870 609from 9203 the 20% most costly BSCs. By BSC population comes 4. Identify cost/benefit for unit replacement Bélgica +32 53 73 42 41 Italia +39 02 95059 554 USA/Canadá +1 866 984 3766 targeting the 20% most costly BSCs at an institution and China +86 21 6865 4588 or Japónreplacement +81 45 453 9220 5. Identify cost/benefit for timed unit +86 10 8419 3588 Nueva Zelanda +64 9 980 6700 Otrosthe países de and Asia +852 reviewing use need2885 for4613 external exhaust, the energy España/Portugal +34 93 223 09 18 Nórdicos/Bálticos/CIS Otros paises +49 6184 90 6000 efficiency of those high use units and the availability of Figure 5 below ranks our surveyedPaises BSCs in order of +358 9 329 10200 o +33 2 2803 2180 more advanced features to reduce consumption on high use annual operating cost from least costly (on the left) to units, then significant savings can be achieved. most costly (on the right). - service history (age, last filter change, last motor SPANISH speed setting) change, current thermoscientific.com