Compliance with BS3978/ISO 3696 Specification for
Transcription
Compliance with BS3978/ISO 3696 Specification for
PURELAB Option-R — Compliance with BS3978/ISO 3696 Specification for “Water for laboratory use”, Grade 1 Introduction The PURELAB Option-R is a self-contained water purifier fed with mains water. It produces highly purified water for general laboratory use, suitable for most analytical, clinical, pharmaceutical and molecular biological applications. Its product water conforms to BS3978/ISO 3696 Grade 1 specification Results Parameter Resistivity (MΩ-cm at 25ºC) Silica (ppb) Absorbance at 254nm Grade 1 specification Option-R specification Typical measured values > 10 > 10 > 15 < 500 <1 < 0.001/cm < 0.001 Method A PURELAB Option-R 15 with boost pump was fed with mains water with a conductivity of 605µS/cm, hardness of 300ppm calcium carbonate and a TOC of 0.7ppmC. The unit was operated intermittently to simulate typical use patterns. Resistivity was logged on-line using a Thornton 200CR meter. Samples were taken for silica analyses and absorbance measurements at various times. Samples were analysed for silica by ICP-MS using a PlasmaQuad 3. PDS/P-Option08/Dec00/E2