3M™ Ground Man Plus Workstation Monitor CTC334

Transcription

3M™ Ground Man Plus Workstation Monitor CTC334
3M™ Ground Man Plus Workstation
Monitor CTC334
User’s Guide
Table of Contents
Section Page
Safety Information.................................................................................................3
1.0 Intended Use.................................................................................................3
2.0 Environmental Conditions............................................................................7
3.0 Package Contents.........................................................................................8
4.0 Description...................................................................................................8
5.0 Installation..................................................................................................10
6.0 Enabling or Disabling Monitoring of Ground Lines..................................11
7.0 Setting Impedance Alarm Levels...............................................................11
8.0 Verification Check......................................................................................12
9.0 Operations..................................................................................................13
10.0 Accessory Items.........................................................................................14
11.0 Maintenance...............................................................................................14
12.0 Customer and Technical Service................................................................15
13.0 Regulatory Information..............................................................................15
2
1.0 Safety Information
Please read, understand, and follow all safety information contained in these
instructions prior to the use of this device. Retain these instructions for future
reference.
Intended Use
The 3M™ Ground Man Plus Workstation Monitor CTC334 is intended for use
by electrical assembly personnel to monitor ground impedance for process and
equipment tools.
The Ground Man Plus Workstation Monitor CTC334 is both a wrist strap
monitor and a ground monitor. It provides operator grounding, and can be used
at work areas or locations to monitor the ground impedance of process tools
in semiconductors, disk drives, flat panels, and EMS and other environments.
Ground impedance monitoring can be used for up to three points for metal
grounds within a tool. The Ground Man Plus Workstation Monitor CTC334 can
also be used to monitor the resistance to ground and voltage level on a person.
The units incorporate green and red LEDs and audible alarms.
The systems must be installed as specified in the user’s guide in an indoor
commercial/industrial environment, and have not been evaluated for other
uses or locations. If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may be impaired.
Explanation of Signal Word Consequences
WARNING
Indicates a hazardous situation which, if not avoided, could result in
death or serious injury.
CAUTION
Indicates a hazardous situation which, if not avoided, could result in
minor or moderate injury and/or property damage.
NOTICE
Indicates a situation which, if not avoided, could result in property damage.
3
Explanation of Symbols
Ground
Double insulation
Direct Current
Terminal
Caution
For Indoor Use Only
WARNING
To reduce the risks associated with hazardous voltage and fire:
• Do not use the power supply if damaged. Do not modify or attempt to service the power supply.
Contact 3M authorized service center for replacement.
• Replace power supply if damaged using only 3M supplied parts.
• Do not attempt to service the 3M™ Ground Man Plus Workstation Monitor CTC334, there are no
user serviceable parts; Return to 3M for service.
• Do not position the Ground Man Plus Workstation Monitor CTC334 accessories or other equipment
where unplugging the power supply is
difficult.
• Always locate the power source (socket or outlet) near the equipment. The power supply plug
serves as the disconnect device.
• Do not use the Ground Man Plus Workstation Monitor CTC334 or its power supply outdoors or wet/
humid environments.
• Do not use the Ground Man Plus Workstation Monitor CTC334 or its power supply outside of the
operating conditions listed in this user guide.
• Use only a dry cloth when cleaning.
To reduce the risks associated with hazardous voltage:
• Use only the power supply provided by 3M and specified for the country of use.
• Before installing the Ground Man Plus Workstation Monitor CTC334 for use with any equipment,
make sure that the Monitor is properly grounded.
To reduce the risks associated with medical device malfunction:
• Persons with heart pacemaker devices should never use this Ground Man Plus Workstation Monitor
CTC334.
CAUTION
To reduce the risks associated with environmental contamination:
• Dispose of Monitor and Power Supply in accordance with all applicable local and government
regulations.
4
1.0 Consignes de sécurité
Veuillez lire, comprendre et suivre toutes les consignes de sécurité figurant dans
les présentes instructions avant d’utiliser ce dispositif. Conserver ces instructions
afin de pouvoir les consulter ultérieurement.
Usage prévu
Le contrôleur pour station de travail Ground Man Plus de 3M™ est conçu pour
être utilisé par le personnel de montage électrique et pour surveiller l’impédance
de terre des outils de processus et d’équipement.
Le contrôleur pour station de travail Ground Man Plus est à la fois un moniteur
de bracelets et un moniteur de terre. Il assure la liaison à la terre de l’opérateur
et peut être utilisé dans les salles ou stations de travail afin de surveiller
l’impédance de terre des outils de processus pour semi-conducteurs, lecteurs
de disques durs, écrans plats, SEM et autres environnements. La surveillance
de l’impédance de terre peut être utilisée pour trois points maximum de masses
métalliques dans un même outil. Le contrôleur pour station de travail Ground
Man Plus peut également être utilisé pour surveiller la résistance à la masse et le
niveau de tension d’une personne. Les unités disposent de DEL vertes et rouges,
ainsi que d’alarmes sonores.
Les systèmes doivent être installés, tel qu’indiqué dans le guide de l’utilisateur,
à l’intérieur d’un environnement commercial/industriel. Les autres usages ou
environnements n’ont pas été évalués. Si l’équipement est utilisé d’une manière
non spécifiée par le fabricant, cela peut compromettre la protection assurée par
l’équipement.
Explication des mots de mise en garde et de leurs conséquences
AVERTISSEMENT
Indique une situation dangereuse, qui si elle n’est pas évitée, peut
entraîner la mort ou des blessures graves.
ATTENTION
Indique une situation dangereuse, qui si elle n’est pas évitée,
peut entraîner des blessures légères ou de gravité modérée et/ou
endommager l’équipement.
MISE EN GARDE
Indique une situation, qui si elle n’est pas évitée, peut endommager
l’équipement.
5
Explication des symboles
Mise à la terre
Double isolation
Courant direct
Borne
Attention
Réservé à un usage en intérieur
AVERTISSEMENT
Pour réduire les risques associés aux incendies et tensions dangereuses :
• Ne pas utiliser l’alimentation électrique si elle est endommagée. Ne pas modifier ou tenter
de réparer l’alimentation électrique. Contacter le centre de réparation agréé 3M pour la faire
remplacer.
• Remplacer l’alimentation électrique en cas de dommage à l’aide de composants fournis par 3M
uniquement.
• Ne pas tenter de réparer le contrôleur pour station de travail Ground Man 3M™, car il ne comporte
aucune pièce réparable par l’utilisateur ; renvoyer à 3M pour réparation.
• Ne pas placer les accessoires du contrôleur pour station de travail Ground Man Plus ou d’autres
équipements susceptibles de gêner le débranchement de l’alimentation électrique.
• Toujours placer l’alimentation électrique (prise ou fiche) près de l’équipement. La fiche de
l’alimentation électrique permet de déconnecter le dispositif.
• Ne pas utiliser le contrôleur de station de travail Ground Man Plus ou son alimentation électrique à
l’extérieur ou dans les environnements humides.
• Ne pas utiliser le contrôleur de station de travail Ground Man Plus ou son alimentation électrique en
dehors des conditions d’exploitations figurant dans le présent guide d’utilisation.
• Nettoyer le système à l’aide d’un chiffon sec.
Pour réduire les risques associés aux tensions dangereuses :
• Utiliser uniquement l’alimentation électrique fournie par 3M et spécifiée pour le pays
correspondant.
• Avant d’installer le contrôleur de station de travail Ground Man Plus en vue de son utilisation avec
d’autres équipements, vérifier que le contrôleur est correctement mis à la terre.
Pour réduire les risques associés aux dysfonctionnements de dispositifs médicaux :
• Les personnes portant un stimulateur cardiaque ne doivent jamais utiliser le contrôleur de station
de travail Ground Man Plus.
ATTENTION
Pour réduire les risques associés aux contaminations environnementales :
• Mettre le contrôleur et son alimentation électrique au rebut conformément à l’ensemble des
réglementations locales et gouvernementales en vigueur.
6
NOTICE
To reduce the risk of damage to components or assemblies being handled:
• 3M™ Ground Man Plus Workstation Monitor CTC334 must be checked periodically to verify each test
mode is functioning correctly.
• Ensure proper operation of Ground Man Plus Workstation Monitor CTC334 by performing operational
verification test as required.
• Always properly ground your tools, equipment, and dissipative surfaces to known good ground
before connecting Ground Man Plus Workstation Monitor CTC334s for monitoring.
MISE EN GARDE
Pour réduire le risque d’endommagement des composants ou ensembles manipulés :
• Le contrôleur de station de travail Ground Man Plus de 3M™ doit être régulièrement contrôlé, afin de
vérifier que chaque mode de test fonctionne correctement.
• Assurer le bon fonctionnement du contrôleur de station de travail Ground Man Plus en effectuant les
tests de vérification du fonctionnement requis.
• Toujours correctement relier vos outils, équipements et surfaces dissipatives à une prise de terre
fonctionnelle avant de brancher le contrôleur de station de travail Ground Man Plus et de lancer la
surveillance.
2.0 Environmental Conditions
This equipment has been tested and found to be safe to operate within these
environmental conditions:
• This is not a warranty of equipment performance within these conditions.
• Indoor use only
• Ingress Protection: IPX0
• Altitude: Up to 2,000 m
• Mains supply voltage fluctuations up to ± 10% of the nominal voltage.
• Transient overvoltages up to the levels of overvoltage category II.
• Temporary overvoltages occurring on mains supply.
• Pollution degree 2.
• Temperature: Maximum 110°F / 43°C Minimum 50°F / 10°C
• Humidity: Maximum relative humidity 80% for temperatures up to 31°C
decreasing linearly to 50% relative humidity at 40°C .
7
3.0 Package Contents
Your 3M™ Ground Man Plus Workstation Monitor CTC334 package should
include the following:
3M™ Ground Man Plus Workstation Monitor CTC334 unit
1 ea.
AC/DC Power Adapter
1 ea.
User’s Guide
1 ea.
General Product
If any of the items are missing, please contact 3M, 1-866-722-3736, or one of
3M authorized distributors immediately.
Power Supply
Universal 100 - 240VAC, 0.5A, ~ 50 - 60Hz
Output: 12VDC, 1.5A
Dimension (approx.)
2.4” W x 0.85” H x 2.6” D
(61 mm W x 21 mm H x 66 mm D)
Weight (approx.)
75g
4.0 Description
Overview
The 3M™ Ground Man Plus Workstation Monitor CTC334 can be used at work
areas or locations to monitor the ground impedance of process tools including,
but not limited to semiconductor, disk drive and flat panel environments. Ground
impedance monitoring can be used for up to three points for metal ground
within a tool. The Ground Man Plus Workstation Monitor CTC334 can also be
used to monitor the resistance to ground and voltage level on a person. The unit
incorporates green and red LEDs and audible alarms.
Ground Impedance Monitor
Continuously monitors for ground impedance on three separate channels that are
connected to three process tools. The alarm level is set the same for all channels.
Output Signal
An output signal is available for connecting to the tool’s computer or PLC. The
5V signal is either logic 1 or 0 for indicating the monitored ground resistance/
impedance conditions and for the wrist strap.
8
Voltage on Person Wrist Strap Monitor
The Ground Man Plus Workstation Monitor CTC334 has a feature that allows
continuous monitoring of voltage on a person through a dual conductor wrist
band assembly.
Performance
Impedance Alarm Level
The alarm level can be preset from 1 to 20 ohms (from 1 to 10 ohms in
increments of 1 ohm and from 10 to 20 ohms in increments of 2 ohms) by
selecting through a combination of LED patterns indicated on the front on the
unit. All three monitoring lines are set to the same selected impedance value.
If the measured impedance on a line is lower than the setting, a green LED
will remain illuminated for that line(s). If the impedance is higher than the
preset level, a red LED and audible alarm are activated for that line(s). If the
impedance level remains above the preset alarm level, the red LED and audible
sound remains on.
The 3M™ Ground Man Plus Workstation Monitor CTC334 only monitors proper
connection to ground. The monitored equipment must be properly ground by
itself. A wire must be attached from the Ground Man Plus Workstation Monitor
CTC334 (GND terminals) to approved ground points to monitor grounding of
process tools correctly.
Output Signal Level
The output connector provides a logic 1 (+5V) when all grounds are within
limits and the unit is powered on. If any of the monitored grounds fail, a wrist
strap fails, or power to the unit is lost, the output is logic 0 (0V).
Person Voltage and Resistance Monitor
The alarm level is monitored when a person is wearing a dual conductor wrist
strap assembly and plugged into the front jack. There is a short hold on the body
voltage alarm so that it is not missed. The unit will respond to either positive or
negative voltage. The alarm is also activated when the wrist band is worn too
loosely and the loop resistance level is exceeded. The red LED indicates an over
the resistance limit without a voltage being generated on the person at the same
time. The green LED indicates proper connection of operator. A green LED plus
blinking red LED will indicate if there is a body voltage generated; operator may
still be connected properly. The green LED is off when a dual conductor ground
cord is not plugged into the jack.
9
Enabling and Disabling Audible Alarm
The audible alarm can be enabled or disabled by momentarily pressing a
recessed miniature push-button Set switch located on the back of the unit while
the monitor is powered on. Use a pin to press the switch.
5.0 Installation
Wire Attachment
The AC adapter should not be connected during the
wire attachment process. Attach 18 AWG wires to
the unit as described below. Strip vinyl insulation at
each end of the wires to approximately 1/3 in. (8 mm).
Twist the stranded wire on each end before inserting
wire into each connector location.
Attach ground wire to the GND connector on the back of the unit. Insert a small
blade screw driver into the orange slot above. Push inward and insert the stripped
wire fully into the hole in the green connector. Hold wire fully in and release the
screw driver to allow the wire to catch.
If you are planning to use the 3M™ Ground Man Plus Workstation Monitor
CTC334 output to the tool, attach wire to the Out connector on back of unit.
Insert the wire as described above.
Attach the impedance monitor wire(s) to the chosen ground line(s) G1, G2, G3
connector(s) on back of unit. Insert the wire as described above.
AC/DC Adapter Connection
Connect the circular plug of the AC/DC Adapter into the power jack on the back
of the unit.
Mounting
Determine the mounting location of the Ground Man Plus Workstation Monitor
CTC334. Attach the unit using one of the following recommended methods:
• Screws
• 3M™ Dual Lock™ Fastener
10
6.0 Enabling or Disabling Monitoring of Ground Lines
Sometimes not all three ground points in the tool need to be monitored. This
procedure will set up the number of ground lines (G1, G2, G3) that will be
used, so that there is no need to put jumpers or shunts on the unused ground
monitoring inputs. The AC Adapter should not be plugged in at this moment.
Attach GND line to known good ground.
Attach monitor line(s) (G1, G2, G3) to ground that has been determined to use.
Do not plug any wires into those ground monitor inputs that will be unused.
Press and hold the Set button using a pin. While the button is pressed inward,
plug in the AC Adapter and then release the Set button.
Wait then for approximately 30 seconds. The 3M™ Ground Man Plus
Workstation Monitor CTC334s will beep during this time. After this time, the
monitors will produce one beep and will automatically disable the lines to which
nothing was connected. This procedure can be repeated whenever a change in
the number of ground lines being monitored is required.
Enabling and Disabling Audible Alarm
The audible alarm can be enabled or disabled by momentarily pressing a
recessed miniature push button Set switch located on the back of the unit while
the monitor is powered on. Use a paper clip to press the switch.
7.0 Setting Impedance Alarm Levels
By default, the Ground Man Plus Workstation Monitor
CTC334 factory set to a 10 ohms alarm level.
If this specification needs to be changed, the following
procedure sets the impedance alarm level for all
channels (G1, G2, G3).
Repeat the steps above in Section 4.0.
Right after the beeps stop; continue pressing the Set button until the impedance
level is matched as shown in the table on the next page.
When the beeping stops, the four left-most LEDs (as shown in the photo on
next page) allow for setting of resistance value. G1 & G2 red LEDs are upper
location and G1 & G2 green LEDs are lower. Select the ground impedance value
by pressing the Set button to advance the resistance limit. Turn the LEDs on or
off as indicated in the table on the next page.
11
Not pressing the Set switch for about seven seconds after selecting causes the
limit value to be saved and normal operation started.
Using LED’s on Front of Unit to Set Resistance Alarm Level
Resistance Ohms
LED
Resistance Ohms
1
9
2
10
3
11
4
12
5
14
6
16
7
20
LED
8
8.0 Verification Check
Verification of G1, G2 and G3 monitoring.
1. Prepare the 3M™ Workstation Monitor Checker CTE701A.
2. Disconnect the wire connected to G1, G2 and G3
3. Connect the ground input of the checker to the ground connector of the 3M™
Ground Man Plus Workstation Monitor CTC334. Set the checker to a proper
resistance alarm level (see the Workstation Monitor Checker CTE701A User’s
Guide).
4. Connect the red wire tip of the checker to the device terminal (G1, G2 and
G3)
5. Press the Metal Ground Fail button on the checker. The red LED of either G1,
G2 or G3 should remain on.
6. Press the Metal Ground Pass button on the checker. The green LED of either
G1,G2 or G3 should turn green.
7. Repeat these steps for all enabled ground monitoring inputs.
12
9.0 Operations
Ground Impedance Monitor
Connect the ground wire to an electrical known good ground point.
Connect the output wire to the tool, if you intend to use it; otherwise leave it
open.
Connect the ground monitor inputs to identified locations in a process tool or
several pieces of other equipment. Make sure to connect the monitoring inputs to
different physical locations than those points used for their grounding.
Person Voltage and Resistance Monitor
This is a feature of the 3M™ Ground Man Plus Workstation Monitor CTC334.
Plug the dual conductor ground cord of the wrist strap into the jack on the
front of the unit. If the wrist strap is worn properly, a green LED will come on;
otherwise the red LED will be on. In order to provide good ground connection,
the wrist strap must be worn tightly and the green light on the Ground Man Plus
Workstation Monitor CTC334 is an indication of that.
If, for some reason, while properly connected, the operator develops excessive
body voltage (in excess of ± 2.5V in any polarity), the green LED may still be
on, but the red LED will blink and the alarm sound will be on. No LED is on if a
dual conductor ground cord is not plugged into the jack.
13
Connecting the 3M™ Ground Man Plus Workstation Monitor CTC334
Facility ground bar
or other good
Output can go to the PLC of the tool
to stop it when any of the grounds
fail or to turn on the light tower
Up to three ground monitoring points
(G1, G2, G3) can monitor the different
parts of the tool
10.0 Accessory Items
3M Product No.
Description
2368
3M™ Dual Conductor Fabric Wrist Band, Adjustable
2360
3M™ Dual Conductor Ground Cord, Coiled, 5 ft.
2370
3M™ Dual Conductor Ground Cord, Coiled, 10 ft.
CTE701A
3M™ Workstation Monitor Checker
11.0 Maintenance
Procedures for Cleaning and Decontamination
Unplug power from the device. Clean using a dry brush or lint free cloth around
the device. In case of contact malfunction, clean contacts using a contact cleaner
or a brush and tighten all connections. Plug in the power.
Repairs and Servicing
Do not attempt to repair the product yourself. Contact a 3M sales representative
or authorized dealer to request inspection and repair. Replace power supply if
damaged using only 3M supplied parts. Other than replacement of remote jacks,
do not attempt to service the device. There are no user-serviceable parts.
14
Device Calibration
Contact a 3M sales representative or authorized dealer to request for product
calibration if needed.
12.0 Customer and Technical Service
For more technical information, call the 3M Electronics Materials Solutions
Division Technical Service Help Line: 866-722-3736 or contact the local 3M
Technical Service.
13.0 Regulatory Information
FCC
This equipment has been tested and found to comply with the limits for a Class
A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to
cause harmful interference in which case the user will be required to correct the
interference at his own expense.
WEEE Statement
The following information is only for EU-members States: The mark shown
to the right is in compliance with Waste Electrical and Electronic Equipment
Directive 2002/96/EC (WEEE). The mark indicates the requirement NOT to
dispose the equipment as unsorted municipal waste, but use the return and
collection systems according to local law.
cULus Statement
Meets cULus requirements.
CE Statement
Meets CE (European Confomity) requirements.
ICES Statement
This Class A digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe A est conforme à la NMB-003 du Canada.
15
China RoHS
Electronic Industry Standard of the People’s Republic of
China, SJ/T11363-2006, Requirements for Concentration
Limits for Certain Hazardous Substances in Electronic
Information Products.
This symbol, per Marking for the Control of Pollution Caused
by Electronic Information Products, SJ/T11364-2006, means
that the product or part does contain a substance, as detailed in the chart below,
in excess of the following maximum concentration values in any homogeneous
material: (a) 0.1% (by weight) for lead, mercury, hexavalent chromium,
polybrominated biphenyls or polybrominated diphenyl ethers; or (b) 0.01% (by
weight) for cadmium. Unless otherwise stated by 3M in writing, this information
represents 3M’s best knowledge and belief based upon information provided by
third party suppliers to 3M.
Name and Content of Hazardous Substances or Elements
Hazardous Substances or Elements
Part or Component Name
(Pb)
(Hg)
(Cd)
(Cr (VI)
(PBB)
(PBDE)
Termination in capacitor 0603
X
O
O
O
O
O
Solder in diode
X
O
O
O
O
O
Finish in diode
X
O
O
O
O
O
Terminations in PCBs
X
O
O
O
O
O
Terminations in resistors 0603
X
O
O
O
O
O
Plating in resistors 0603
X
O
O
O
O
O
Resistor ink in potentiometer
X
O
O
O
O
O
Solder in instrument
X
O
O
O
O
O
Solder in IC
X
O
O
O
O
O
Solder in buzzer
X
O
O
O
O
O
Audio jack
X
O
O
O
O
O
O: Indicates that this hazardous substance contained in all of the homogeneous materials for this part is below the limit
requirement in the SJ/T11363-2006.
X: Indicates that this hazardous substance contained in at least one of the homogeneous materials used for this part is above the
limit requirement in the SJ/T11363-2006.
16
ACPEIP
Norme d’industrie électronique de la République populaire
de Chine, SJ/T11363-2006, Exigences relatives aux limites
de concentration de certaines substances dangereuses
contenues dans les produits d’information électronique.
Ce symbole, repère du Contrôle de la pollution causée
par les produits d’information électronique, SJ/T113642006, signifie que le produit ou l’un de ses composants contient une
substance, figurant dans le tableau ci-dessous, dans une quantité excédant les
valeurs de concentration maximales suivantes dans un matériau homogène
: (a) 0,1 % (du poids) pour le plomb, le mercure, le chrome hexavalent, les
polybromobiphényles ou les polybromodiphényléthers ; ou (b) 0,01 % (du poids)
pour le cadmium. Sauf indication écrite contraire de 3M, ces informations sont
fondées sur les meilleures connaissances et hypothèses de 3M, ainsi que sur des
informations fournies par les tiers fournisseurs de 3M.
Nom et composition des substances ou éléments dangereux
Substances ou éléments dangereux
Nom de la pièce ou du composant
(Pb)
(Hg)
(Cd)
(Cr (VI)
(PBB)
(PBDE)
Borne de condensateur 0603
X
O
O
O
O
O
Soudure de la diode
X
O
O
O
O
O
Terminaison de la diode
X
O
O
O
O
O
Bornes en PCB
X
O
O
O
O
O
Bornes de résistances 0603
X
O
O
O
O
O
Revêtement métallique de résistances 0603
X
O
O
O
O
O
Encre de résistance de potentiomètre
X
O
O
O
O
O
Soudure de l’instrument
X
O
O
O
O
O
Soudure de CI
X
O
O
O
O
O
Soudure de vibreur
X
O
O
O
O
O
Prise jack audio
X
O
O
O
O
O
O : indique que la quantité de cette substance dangereuse contenue dans tous les matériaux homogènes de cette
pièce est inférieure aux limites définies dans la SJ/T11363-2006.
X : indique que la quantité de cette substance dangereuse contenue dans tous les matériaux homogènes de cette
pièce est supérieure aux limites définies dans la SJ/T11363-2006.
17
产品中有毒有害物质或元素的名称及含量
有毒有害物质或元素
部件名称
铅
汞
镉
六价铬
多溴联苯
多溴二苯醚
0603电容引脚
X
O
O
O
O
O
二极管焊接部
X
O
O
O
O
O
二极管电镀
X
O
O
O
O
O
印刷电路板焊盘/安装孔
X
O
O
O
O
O
0603 电阻引脚
X
O
O
O
O
O
0603 电阻电镀部
X
O
O
O
O
O
电位器中的电阻油墨
X
O
O
O
O
O
装置的焊接部
X
O
O
O
O
O
IC焊接部
X
O
O
O
O
O
蜂鸣器焊接部
X
O
O
O
O
O
音频插孔
X
O
O
O
O
O
O: 表示该有毒有害物质在该部件所有均质材料中的含量均在SJ/T11363-2006 标准规定的限量要求以下。
X: 表示该有毒有害物质至少在该部件的某一均质材料中的含量超出SJ/T11363-2006 标准规定的限量要求。
18
3M and Dual Lock are a trademarks of 3M Company.
Important Notice
All statements, technical information, and recommendations related to 3M’s products
are based on information believed to be reliable, but the accuracy or completeness is
not guaranteed. Before using this product, you must evaluate it and determine if it is
suitable for your intended application. You assume all risks and liability associated with
such use. Any statements related to the product which are not contained in 3M’s current
publications, or any contrary statements contained on your purchase order shall have no
force or effect unless expressly agreed upon, in writing, by an authorized officer of 3M.
Warranty; Limited Remedy; Limited Liability.
This product will be free from defects in material and manufacture for one year from
the time of purchase. 3M MAKES NO OTHER WARRANTIES INCLUDING, BUT
NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE. If this product is defective within
the warranty period stated above, your exclusive remedy shall be, at 3M’s option, to
replace or repair the 3M product or refund the purchase price of the 3M product. Except
where prohibited by law, 3M will not be liable for any indirect, special, incidental or
consequential loss or damage arising from this 3M product, regardless of the legal
theory asserted.
3