Release 4.2 Service-Pack 1 For the content of this service pack, see
Transcription
Release 4.2 Service-Pack 1 For the content of this service pack, see
Release 4.2 Service-Pack 1 For the content of this service pack, see the list below. New functions Full support of AxioCam HRm Rev. 3.0 available Full support of AxioCam HRc Rev. 3.0 available Full support of AxioCam MRc Rev. 3.0 available Solved issues of Rel. 4.2 Reuse MultiTrack and Lambda Scan and 4 Channel Track: the settings for one channel are not reused. Support for Systems without “none” filter in NT 3 solved Error message in pinhole adjustment on LSM 5 MP system The modify wavelength dialog for MaiTai-Laser loses the input focus of the laser. VME Abort during time-series blocks execution during the next run of the macro Modification of Ref-Corr-Slider causes unexpected focus movement (Piezo Objective). Bleach ROIs are not listed in Mean ROI-List for LIVE Stripe-Bleach LSM-DuoScan: Its not possible to chose a different ZPosition for bleaching using the DuoScan and lasers from the LIVE laser module and image with LSM 5 LIVE (not functional for simultaneous bleaching and imaging) FCS: a number of crashes during operating, printing and closing the software were solved FCS: Loading of a method does not load the cross correlation fit settings FCS: Not possible to fix diffusion time in single component fits. FCS: Wrong display of Gamma and molecules in some extended fit models. StitchArt: Display of Z values wrong when performing 3D stiching. Display of Z axis in diagram and for the export wrong. solved solved solved solved solved solved solved solved solved solved solved solved