Model 6000 Series CA-CALC Single Gas Monitors

Transcription

Model 6000 Series CA-CALC Single Gas Monitors
Combustion Analysis
Model 6000 Series
CA-CALCTM
Single Gas Monitors
Operation and Service Manual
P/N 1980452 Revision B
November 2003
Model 6000 Series
CA-CALCTM
Single Gas Monitors
Operation and Service
Manual
November 2003
P/N 1980452, Revision B
U.S. and CANADA
Sales and Customer Service:
(800) 874-2811/(651) 490-2811
Fax:
(651) 490-3824
MAIL/SHIP TO:
TSI Incorporated
ATTN: Customer Service
500 Cardigan Road
Shoreview, MN 55126
USA
INTERNATIONAL
Sales and Customer Service:
(001 651) 490-2811
Fax:
(001 651) 490-3824
E-Mail
[email protected]
WEB SITE
www.tsi.com
Copyright ©
TSI Incorporated / 2003 / All rights reserved.
Part Number
1980452 / Revision B
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective July 2000)
Seller warrants the goods sold hereunder, under normal use and service as described in the
operator's manual, shall be free from defects in workmanship and material for twenty-four (24)
months, or the length of time specified in the operator's manual, from the date of shipment to the
customer. This warranty period is inclusive of any statutory warranty. This limited warranty is
subject to the following exclusions:
a.
Batteries are warranted for 90 days from the date of shipment to the customer.
Electrochemical gas sensors are warranted for a period of twelve (12) months from the date
of shipment to the customer.
b. Parts repaired or replaced as a result of repair services are warranted to be free from defects
in workmanship and material, under normal use, for 90 days from the date of shipment.
c. Seller does not provide any warranty on finished goods manufactured by others or on any
fuses, batteries or other consumable materials. Only the original manufacturer's warranty
applies.
d. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with
respect to, and shall have no liability in connection with, goods which are incorporated into
other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated
herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR
BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED
ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL
BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE
PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE
GOODS. IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL,
CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE
RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS
OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12
months after a cause of action has accrued. The goods returned under warranty to Seller's
factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND
LIABILITY, which contains the complete and exclusive limited warranty of Seller. This
LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms
waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our
customers, our service policy is designed to give prompt attention to any problems. If any
malfunction is discovered, please contact your nearest sales office or representative, or call TSI's
Customer Service department at (800) 874-2811.
CONTENTS
Chapters
Introduction ........................................................................................ 1
Manual Purpose ............................................................................ 1
Using This Manual......................................................................... 1
Warnings and Cautions ................................................................. 1
1. Instrument Description ............................................................... 3
2. Unpacking .................................................................................... 5
List of Standard Components........................................................ 5
Accessories and Replacement Parts ............................................ 5
3. Component Identification ........................................................... 7
The Gas Sensors .......................................................................... 9
The Sampling Probe...................................................................... 9
Diaphragm Pump .......................................................................... 9
Water-Stop Filter ........................................................................... 9
Optional Model 802215 Water Trap .............................................. 9
Optional Protective Boot Model 801995...................................... 10
4. Getting Started .......................................................................... 11
Assembling the Sampling Probe ................................................. 11
5. Basic Operation ......................................................................... 13
Buttons and Button Operations ................................................... 13
Startup ......................................................................................... 13
Connecting the Sampling Probe ................................................. 13
Data Display, Viewing the Gas Measurement ............................ 15
Changing the Gas Measurement Units (not CA-6020) ............... 15
Making Gas Measurements ........................................................ 16
6. Gas Sensor Calibrations........................................................... 17
Calibration Equipment Setups..................................................... 17
CAL MENU Option, Sensor Calibration ...................................... 19
7. Maintenance and Troubleshooting.......................................... 21
Water-Stop Filter Plugging .......................................................... 21
Changing the Water-Stop Filter................................................... 21
Cleaning the Sample Probe ........................................................ 21
Adding or Replacing a Gas Sensor............................................. 22
CAL.F MENU Option—Setting the Calibration Factor for a
New Sensor.............................................................................. 24
i
Appendixes
A. Error Codes Defined..................................................................25
B. Series CA-6000 CA-CALC Single Gas Monitor Detailed
Specifications ............................................................................27
ii
Introduction
Manual Purpose
This manual describes the operation and maintenance of your TSI Series
CA-6000 CA-CALCTM Single Gas Monitor.
Using This Manual
Before using the CA-CALC Single Gas Monitor for the first time, review this
manual in its entirety.
Most information in this manual is applicable for all Series CA-6000
instruments. Where information is not applicable to a particular model or
additional information is supplied, notation is made.
Warnings and Cautions
The manual assumes that you have a basic understanding of combustion
safety concerns and are thoroughly familiar with the fuel burning equipment
being tested. If you are using measurements as the basis for equipment
adjustments, rely on your good judgment and experience together with the
measured data. This is especially important where safety issues are of
concern. Equipment adjustments must always coincide with the fuel burning
equipment manufacturer’s recommendations.
!
!
WARNING
High temperatures and toxic gases are produced when
fossil fuels are burned. Only qualified individuals,
thoroughly familiar with operating and adjusting fuelburning equipment, should use gas measurements for
the purpose of making equipment adjustments.
WARNING
The CA-CALC is not intended for use as a continuous
monitor.
1
Chapter 1
Instrument Description
The CA-CALCTM 6000 Series Single Gas Monitors are portable instruments
designed for measuring key gases associated with the combustion of fossil
fuels. The CA-CALC Single Gas Monitor is typically used to evaluate the
performance of boilers, furnaces, and hot water tanks, and determine the
presence of toxic pollutants for regulatory and safety concerns.
All instruments use electrochemical gas sensors, which for most models
provide two or three years of useful service before sensor replacement is
necessary. Sensor replacement is easily performed by the operator.
The CA-CALC 6000 Series Analyzers are supplied with a sampling probe
consisting of a stainless sample tube, plastic handle, rubber or non-reactive
plastic sample line (for NO2, SO2), and in-line water-stop filter. The
sampling probe is used to draw the gas sample at a rate of 0.8 liters per
minute from the sampling point to the sensor in the instrument. The sample
flow is provided by an internal diaphragm vacuum pump.
The following CA-CALC models are available:
Model Number
CA-6010
CA-6020
CA-6030
CA-6040
CA-6050
CA-6060
CA-6070
Gas Measurement
CO, carbon monoxide 0 – 1000 PPM
O2, oxygen 0 – 25%
NO, nitric oxide 0 – 4000 PPM
NO2, nitrogen dioxide 0 – 500 PPM
SO2, sulfur dioxide 0 – 4000 PPM
CO (high) high concentration CO 0 – 2%
CO (X-high) high concentration CO 0.1% – 8%
3
Chapter 2
Unpacking
Carefully unpack your CA-CALCTM Single Gas Monitor and accessories
from the carrying case. Check the individual parts against the list of
components in the table below. If items are missing or damaged, notify
TSI immediately.
List of Standard Components
Qty.
Item Component
1
Series CA-6000 CA-CALC single gas monitor
1
Sample probe with water-stop filter
801993
1
Sample probe with water-stop filter, for
Models CA-6040, CA-6050
AA cell alkaline batteries
Operation and Service manual
801998
4
1
Part/ Model
1980452
Accessories and Replacement Parts
Item
Part/Model
For Model
Boot with magnet and carrying strap
Water-stop filter replacement kit
Optional Water trap
Water trap filters for optional Water Trap
Lithium battery (Model 6030)
CO replacement sensor
O2 replacement sensor
NO replacement sensor
NO2 replacement sensor
SO2 replacement sensor
CO high concentration replacement sensor
CO x-high replacement sensor to 8%
Replacement pump
O2 sensor zero calibration kit (N2 gas)
CO sensor calibration kit, 500 ppm CO
NO sensor calibration kit, 200 ppm NO
NO2 sensor calibration kit, 25 ppm NO2
SO2 sensor calibration kit, 200 ppm SO2
CO high concentration calibration kit, 1%
801995
801922
802215
1602309
1208023
802006
802012
802007
802008
802009
802010
802013
801996
801939
801999
801937
801938
801936
801945
All
All
All
All
CA-6030
CA-6010
CA-6020
CA-6030
CA-6040
CA-6050
CA-6060
CA-6070
All
CA-6020
CA-6010
CA-6030
CA-6040
CA-6050
CA-6060/70
5
6
Chapter 3
Chapter 3
Component Identification
Key components of the CA-CALCTM Single Gas Monitor and sampling
probe are identified in Figures 1 and 2 and under section headings in the
text that follows.
1. Buttons
2. Display
3. Top cover
4. Case bottom
5. Battery cover
6. Sensor cover
7. Instrument label
8. Gas vent
9. Sample port
Figure 1: CA-CALC CA- 6000 Series
7
6
5
3
1
7
Sampling Probe for Models CA-6010,
CA-6020, CA-6030, CA-6060
8
2
1
4
5
4
6
7
Sampling Probe for Models CA-6040, CA-0650
1. Stainless steel sampling tube
2. Plastic handle
3. Rubber sampling tube
4. Sampling tube (NO2, SO2)
5. Water-stop filter
6. Connection tubing
7. Tube fitting
8. Adapter tube
Figure 2: CA-CALC Sampling Probe Components
8
Chapter 3
The Gas Sensors
Your CA-CALC analyzer uses electrochemical gas sensors for making gas
measurements. Electrochemical sensors operate by oxidizing or reducing
molecules of the measured gas contacting the sensor’s sensing electrode.
A sensor current develops which is proportional to the reaction rate. The
current is detected by the CA-CALC electronics and converted to a gas
concentration reading proportional to the current. Gas concentration is
displayed in units of PPM or standard mg/m3, or for oxygen (O2),
displayed as a percent.
The Sampling Probe
The sampling probe consists of a stainless steel probe tip, plastic handle,
sampling tube and water-stop filter. Instrument Models CA-6040 and
CA-6050 for measuring NO2 and SO2, respectively, use a polypropylene
sample tube, which is non-reactive to these gases. Sampling probe types
are shown in Figure 2.
Diaphragm Pump
The CA-CALC analyzer samples exhaust gases from the flue and delivers
them to the electrochemical sensors using a diaphragm sampling pump.
The pump is accessed through the sensor cover (see Figure 1), and can be
removed for cleaning or for replacement. Typical pump life is 1000
hours.
Water-Stop Filter
Your sample probe is provided with a small, water-impermeable
hydrophobic filter. This filter protects the pump and gas sensor from
water (and soot), which can adversely affect instrument performance.
When water is present, the filter plugs, causing the instrument flow to
cease. Where water or contamination are present (flue measurements), it
is recommended that you purchase a Model 802215 Water Trap. This
item is described below.
Optional Model 802215 Water Trap
The Model 802215 Water Trap, shown in Figure 2, is used to remove
moisture that collects in the sample tubing when combustion gases are
sampled from the flue. The water trap uses two chambers and a filter to
maximize water and particulate removal. Unlike the water-stop filter
provided standard, the Water Trap enables extended sampling from a
moist, dirty air stream.
Component Identification
9
Optional Protective Boot Model 801995
An optional protective boot is available for your instrument. The
boot provides drop protection for the instrument and is supplied
with a magnet, enabling the instrument to be mounted on a flat,
vertical metal surface. The boot is equipped with a carrying strap.
10
Chapter 3
Chapter 4
Getting Started
Installing Batteries
The CA-CALCTM portable Single Gas Monitor operates using 4 AA batteries
enabling the instrument to operate for approximately 10 hours with the pump
running.
To install the batteries turn the Single Gas Monitor over and remove the
battery cover by pushing down on the cover latch tab with your thumb and
sliding back, away from the instrument case. Turn the instrument over and
shake, allowing the battery holder and batteries to fall out into your hand.
Note the orientation of the contacts on the holder and in the instrument.
Replace the used batteries with new batteries, matching the + and –
terminals, as indicated on the battery holder. Note the location of the battery
contacts. Replace the holder by inserting it straight down into the battery
compartment
Assembling the Sampling Probe
Refer to the photographs in Figure 2. Your probe comes assembled as
shown. Should the water-stop filter become separated, reconnect it by
attaching the short black rubber tube (1/8” ID) to the smooth “Lure” fitting.
Connect the larger black rubber tube (3/16” ID) to the barbed tube fitting
attached to the water-stop filter. Where the transparent polypropylene
tubing is supplied (probes for NO2 and SO2), reconnect the poly-tubing by
sliding it into the rubber tubing as shown in the photos.
11
Chapter 5
Basic Operation
Buttons and Button Operations
ON-OFF Control Button
Turns the instrument on and off.
The ENTER Control Button
Press the ENTER button to execute a command,
such as selecting a menu item.
The ESC Control Button
Used to return to the previous menu or cancel a
process.
ARROW Control Buttons
Use the arrow buttons to step between items and
change values.
The PUMP On-Off Button
Turns the pump on or off.
Startup
Remove the sampling probe from the flue or disconnect the sample tubing
from the sampling port. Press the ON-OFF button. Figure 3 diagrams the
start-up sequence and shows screens displayed during startup. If no errors
are detected, the Data Display screen appears when the sequence is
complete.
While the WARM screen is displayed, the O2 sensor is calibrated using
ambient air, and other gas sensors are zeroed.
Connecting the Sampling Probe
The sampling probe is connected to the instrument by pushing the short
rubber sample tube completely over the instrument sample port.
13
Figure 3. Startup Sequence
14
Chapter 5
Data Display, Viewing the Gas Measurement
Measurements and calculations are presented in the Data Display screen.
The Data Display appears once the Startup sequence is complete. Except
for the Model 6020 (O2), gas concentrations are presented in PPM (parts per
million) or mg/m3 (milligrams per cubic meter). Refer to the section below
for instructions on changing the units of measurement.
Changing the Gas Measurement Units (not CA-6020, 6070)
Units of gas concentration are changeable. Parts per million, PPM, or
milligrams per standard cubic meters, mg/m3 or percent (CA-6060) can be
presented on your instrument Data Display. Note: mg/m3 is referenced to 0
deg. C and 760 mm Hg.
To change units, press the ENTER button. Use the ARROW buttons to and
scroll to UNITS. Press ENTER. Select either PPM or mg/m3. Press
ENTER. The figure below outlines steps for changing units.
Figure 4. Choosing Display Units
Basic Operation
15
Making Gas Measurements
Your instrument is ergonomically designed to be held in the left hand. The
sampling probe is held in the right, or may be mounted. If you have the
protective boot option Model 801995, an attached magnet enables the
instrument to be mounted vertically to the metal side of your appliance, for
hands-free operation.
Measurement Interval
When making gas measurements, keep the measurements short.
Although the instrument is capable of measurements lasting many
minutes, the gas sensors are designed for discrete sampling applications.
When high gas concentrations are measured for extended periods,
chemical filters in the sensor are used up prematurely. The chemical
filters remove interfering gases that can cause false readings,
influencing the accuracy and lifetime of the sensor. In addition, soot
and moisture will plug the water-stop filter, reducing or stopping the gas
flow.
Note: To reduce sensor exposure to gas and to reduce build up of water
vapor in the sampling lines and water trap, turn the pump off
when not making measurements.
Operating Temperature
Best results are obtained if the CA-CALC single gas monitor is allowed
to stabilize at the temperature of the test environment before using.
Plugged Water-Stop Filter
A significant change in the sound of the pump may indicate a plugged
water-stop filter. If plugging is suspected, refer to the appropriate
heading in the Maintenance section of this manual.
High Temperature Measurements
When sampling from high temperature, leave at least four inches
between the plastic handle of the probe and the outside edge of the flue.
The stainless-steel sampling probe is a poor heat conductor, and
following the four-inch rule above, temperatures as high as 700 degrees
centigrade (1290ºF) can be measured without damaging the probe
handle.
16
Chapter 5
Chapter 6
Gas Sensor Calibrations
Calibration Equipment Setups
Gas sensors can be calibrated periodically to maintain the best accuracy.
Gas sensors do drift over time, depending upon the operating environment
and gas sensor’s exposure history.
With the proper equipment, such as that shown in Figures 5 and 6 below, it
is easy to calibrate your CA-CALC gas sensor. If you wish, you may also
return your instrument to TSI for a new factory calibration, or purchase a
new factory calibrated sensor.
The equipment needed to calibrate individual gas sensors can be purchased
from TSI as calibration kits, or you may elect to put together your own
calibration system. Model numbers for TSI sensor calibration kits are found
in Chapter 2, “Unpacking.” Two calibration setups are presented in
Figures 5 and 6. A brief discussion of these calibration setups is presented in
the following section.
Figure 5. Calibration with TSI Calibration Kit
17
A TSI supplied calibration kit (Figure 5) uses a Demand Flow regulator to
supply gas to the CA-CALC analyzer in response to the draw of the
instrument sampling the pump. If a conventional regulator and valve are
used (Figure 6), the setup supplies gas to the instrument using a tee to a vent
extra gas. This prevents a forced flow at the instrument inlet. The bead-type
flow meter depicted in the figure is used to verify there is extra flow (.5 to
2 LPM recommended). Extra flow is required to prevent room air from
being drawn in, diluting the sample.
Once you have the assembled the appropriate hardware described, follow the
steps in the next section to perform the gas sensor calibration.
Figure 6. Alternative Calibration Setup
18
Chapter 6
CAL MENU Option, Sensor Calibration
The CAL option is used for performing calibration of the gas sensor. A
schematic showing the button operations and screen displays for sensor
calibration is shown in Figure 7. The calibration procedures are also
described below.
1. To begin, press the ENTER button from the Data Display screen. The
CAL screen appears. CAL is one of three MENU options selectable
using the ARROW buttons (reference Figure 4).
2. Press the ENTER button when CAL is displayed. The word ZERO
indicates that the sensor voltage is first read, without calibration gas.
3. Unless calibrating an O2 sensor (Model 6120), disconnect the sampling
line so only room air is sampled. If you are calibrating an O2 sensor,
connect your N2 (nitrogen) calibration bottle to the sampling probe
(refer to Figures 5 or 6).
4. Press ENTER to start the ZERO calibration. Wait for the countdown to
finish.
5. When the zero calibration is complete, SPAN is indicated on the
display. SPAN indicates that the voltage will be measured in the
presence of your calibration gas. For O2, this means the surrounding
ambient air.
6. Connect your calibration gas to the sampling probe as shown in
Figures 5, and 6 described earlier. If calibrating an O2 sensor,
disconnect the N2 gas bottle.
7. Use the ARROW keys to adjust the SPAN value to match that of your
calibration gas. For O2 this is 20.9%, the ambient air concentration.
8. Press ENTER to start the SPAN calibration. Wait for the countdown to
complete.
9. If an error code appears, the initial calibration value will be reinstalled.
Refer to Appendix A.
Notes: „ Pressing the ESC button will bypass either the ZERO or SPAN
calibration.
„ When the calibration is complete, the calibration factor
automatically changes to reflect the new calibration. For more on
the calibration factor refer to the CAL.F section in the
Maintenance chapter.
Gas Sensor Calibrations
19
Figure 7. Gas Calibration Button Sequence. Example, CO
Gas Monitor Model 6010.
20
Chapter 6
Chapter 7
Maintenance and Troubleshooting
Water-Stop Filter Plugging
When water comes in contact with the water-stop filter (Figure 2), it reduces
or stops flow to the gas sensor. If this occurs, the filter needs to be dried out
or replaced.
Check for a plugged flow by blocking the end of the probe with your thumb.
Listen for a lowering of the pump frequency. If no change in tone occurs,
your filter is probably already plugged. Disconnect the sampling probe from
the instrument. The pump sound should change, increasing in frequency.
Change the filter, setting the plugged filter aside to dry. If filter plugging is
a continuing problem, a TSI Model 801997 Water Trap is recommended.
Changing the Water-Stop Filter
1.
2.
3.
4.
5.
Refer to Figure 2.
Remove the sample tubing from the instrument sample port.
Remove this tubing from the filter and save.
Twist the fitting on the end of the filter counterclockwise to separate the
filter from the fitting. Set the filter aside to dry, or discard if clearly
dirty.
Replace with a new filter. Connect the fitting and replacing the short
tubing.
Cleaning the Sample Probe
Cleaning may be necessary in high-soot environments. Soot accumulates in
the steel sampling tube and sampling line and over time may contribute to a
blocked flow path. It is assumed that an optional water trap is present. The
Water-stop filter supplied standard will not perform in environments where
high levels of particulate material (soot) are present.
1.
Disconnect the sampling probe from the instrument.
2.
Remove the water-stop filter or remove the optional Water Trap (if
installed) by separating the sample tube from the tube stub on each end
of the trap.
3.
Rinse the tubing, allowing the water to drain from the probe end. When
the water is clear, discontinue the rinse. Orient the probe and tubing so
excess water drains from the sample lines.
21
4.
Allow adequate time for the interior of the probe to dry. Replace the
water trap making sure the water trap is oriented properly. The filter
must be toward the instrument.
Adding or Replacing a Gas Sensor
Changing the CO Sensor (Model 6010)
1. Remove the sensor cover shown in Figure 9, by removing the
sensor cover screw (2).
2. Carefully lift the rubber sensor manifold (3) from the sensor and
bend it back. Do not attempt to remove the rubber manifold from
the case!
3. Pull the sensor, together with the foam gasket, out of the
instrument.
4. Remove the gasket from the sensor.
5. Replace the gasket in the rubber sensor manifold (3) by aligning the
notch in the gasket with the tab on the manifold (7).
6. Install the new sensor by aligning the pins on the sensor with those
on the instrument electronics board.
7. Replace the sensor manifold, pushing the foam gasket over the new
CO sensor until it is fully seated. Replace the sensor cover and
cover screw.
8. Find your calibration factor sheet included with your new sensor.
Enter the new calibration factor as described in the section “CAL.F
MENU Option” later in this chapter.
22
Chapter 7
1.
2.
3.
4.
Sensor Cover
Sensor retaining screw
Rubber sensor manifold
O2 sensor position
(Model 6020)
5. Other sensors
6. Dummy sensor plug
7. Manifold Tab for CO gasket alignment
Figure 9. Removing and Replacing a Gas Sensor
Changing Other Sensors
1. Remove the sensor cover shown in the figure above, by removing
the sensor cover screw.
2. Carefully lift the rubber sensor manifold from the sensor(s) and
bend it back. Do not attempt to remove the rubber manifold from
the case!
3. Remove sensor from the instrument. Leaving the dummy sensor
plug in place.
4. Install the new sensor by aligning the pins on the sensor with those
on the instrument electronics board. If you have an oxygen sensor
(Model 6020), install the sensor in position number 4 shown in
Figure 9. All other sensors are installed in position 5.
Note: For the oxygen sensor match the + indication on the sensor
with the + pin receptacle on the electronics board.
5. Replace the sensor manifold, pushing the manifold cup over the
new sensor, until it is fully seated.
Maintenance and Troubleshooting
23
6.
Find your calibration factor sheet included with your new sensor.
Enter the new calibration factor as described in the section “CAL.F
MENU Option” which follows.
CAL.F MENU Option—Setting the Calibration Factor for a New
Sensor
When you install a calibrated replacement sensor from TSI, you must
complete the new sensor installation by entering a new calibration factor.
This factor is determined by a sensor calibration at the factory.
After installing your new sensor, find the calibration factor sheet.
Use the MENU option CAL.F, and enter calibration factor provided. Refer
to the details outlined in the schematic below.
Figure 10. Setting the Calibration Factor
24
Chapter 7
Appendix A
Error Codes Defined
Error code by number:
6. Lithium battery low.
9. Can't save data to EEPROM.
12. Model number checksum error.
16. A/D calibration checksum error.
17. Chem sensor checksum error.
30. O2 zero voltage out of range.
31. CO zero voltage out of range.
32. COhi zero voltage out of range.
33. NO zero voltage out of range.
34. NO2 zero voltage out of range.
35. SO2 zero voltage out of range.
36. Instrument temperature out of range.
37. O2 span voltage out of range.
38. A/D converter calibration factor out of range.
39. A/D converter offset factor out of range.
40. O2 gain factor out of range.
41. CO gain factor out of range.
42. COhi gain factor out of range.
43. NO gain factor out of range.
44. NO2 gain factor out of range.
45. SO2 gain factor out of range.
70. Internal program error—contact TSI.
25
Appendix B
Series CA-6000 CA-CALC Single Gas Monitor
Detailed Specifications*
Carbon Monoxide (CO): CA-6010
Range: ......................................0–2000 ppm (9999 ppm max)
Accuracy: ..................................0–100 ±5 ppm or 10% of reading
100–1000 ±10 ppm or 5% (greater of)
Resolution:................................1 ppm
Response Time:........................<30 seconds to 90% of step change
Oxygen (O2) CA-6020
Range: ......................................0–25%
Accuracy: ..................................±0.3% O2
Resolution:................................0.1% O2
Response Time:........................<30 seconds to 90% of step change
Nitric Oxide CA-6030
Range: ......................................0–4000 ppm (5000 ppm max)
Accuracy: ..................................0–100 ±5 ppm
>100–1000 ±5% of reading
>1000–4000 ±10% of reading
Resolution:................................1 ppm
Response Time:........................<30 seconds to 90% of step change
Nitrogen Dioxide (NO2) CA-6040
Range: ......................................0–500 ppm (1000 ppm max)
Accuracy: ..................................0–100 ±5 ppm
>100–200 ±5% of reading
>200–500 ±10% of reading
Resolution:.................................1 ppm
Response Time:........................<30 seconds to 90% of step change
Sulfur Dioxide (SO2) CA-6050
Range: ......................................0–4000 ppm (5000 ppm max)
Accuracy: ..................................0–200 ±10 ppm
>200–1000 ±5% of reading
>1000–4000 ±10% of reading
Resolution:................................1 ppm
Response Time:........................<30 seconds to 90% of step change
*Specifications are subject to change without notice.
27
Carbon Monoxide High Concentration (CO hi) CA-6060
Range:......................................0–2% ppm
Accuracy:..................................0–500 ±100 ppm
>500–9999 ±5% of reading
>1%–2% ±10% of reading
Resolution: ...............................5 ppm
Response Time: .......................<40 seconds to 90% of step change
Carbon Monoxide High Concentration (CO X-hi) CA-6070
Range:......................................0–8% CO
Accuracy:..................................±0.03% CO
Resolution: ...............................0.1% CO
Response Time: .......................<60 seconds to 90% of step change
External dimensions: ............4 x 7 x 2¼” (10 x 18 x 5.7 cm)
Weight: ....................................1.2 lbs
Power Requirements:
Batteries: ..................................4 AA cell alkaline batteries
Battery life: ...............................>10 hours (pump on)
Note: These specifications assume the instrument is allowed to stabilize at the
operating temperature before being turned on.
28
Appendix B
TSI Incorporated
500 Cardigan Road, Shoreview, MN 55126 U.S.A.
Web: www.tsi.com